Quick Start Guide
Conduit installation
Prevent condensation in any conduit from flowing into the housing by mounting
the flowmeter at a high point in the conduit run. If the flowmeter is mounted at a
low point in the conduit run, the terminal compartment could fill with fluid.
If the conduit originates above the flowmeter, route conduit below the flowmeter
before entry. In some cases a drain seal may need to be installed.
Figure 13. Proper Conduit Installation with Rosemount 8800D
A
A. Conduit Line
Use the following steps to wire the flowmeter:
1. Remove the housing cover on the side marked FIELD TERMINALS.
2. Connect the positive lead to the "+" terminal and the negative lead to the "-"
terminal as shown in
F
OUNDATION
Note
F
Fieldbus terminals are not polarity sensitive.
OUNDATION
3. For HART installations utilizing the pulse output, connect the positive lead to
the "+" terminal of the pulse output and the negative lead to the "-" terminal
of the pulse output as shown in
supply is required for the pulse output. Maximum switching current for the
pulse output is 120 mA.
12
Figure 14
fieldbus installations.
A
for HART installations and
Figure
14. A separate 5 to 30 Vdc power
April 2016
Figure 15
for