WARNING: For vessels with conical or parabolic tops, avoid
mounting the instrument at the centre. (The concavity of the top can
focus echoes into the centre, giving false readings.)
Note:
Under certain circumstances, it may be acceptable to mount SITRANS LR 200 at
the centre of a flat-topped tank: please discuss this with your Siemens Milltronics
Representative.
preferred
undesirable
Keep the emission cone free of interference:
Make allowance for the emission cone spreading: allow a minimum of 300 mm (1 ft)
for every 3 m (10 ft) of vessel height.
Locate the antenna away from the side wall, to avoid interference from indirect
echoes.
Avoid interference (from objects such as ladders or pipes, or from the fill path)
which can cause false reflections
Note:
• Beam angle defined at –3dB
boundary
horn antenna
beam angle:
o
30
4" horn
o
20
6" horn
o
12
8" horn
1.
For more detail on False Echo suppression, see
TVT (Auto False Echo Suppression) setup
7ML19985HP01
SITRANS LR 200 (PROFIBUS PA) – INSTRUCTION MANUAL
Coni
Conical
Flat
1
.
Auto False-Echo Suppression
on page 100.
Fl
at
Parabolic
minimum of 300 mm (1ft) for
every 3 m (10') of vessel height.
rod antenna
beam
angle
o
20
on page 35, and