Alarm Text
TELE SYNC UNSUPP
Source - Telemetry
TELE UNSUPPORTED
Source - Patient Monitor
TELE WEAK SIGNAL
Source - Telemetry
TRANSMITTER MALF
Source - Telemetry
TRANSMITTER OFF
Source - Telemetry
Draft - 1 Aug 08
Technical Alarms (Inoperative Conditions)
Priority
Condition
Hard
Incompatible MMS
Hard
The transceiver is not
supported for direct
connection to the monitor.
Soft
Patient is at outer
range of the radio
coverage area.
Telemetry pack is
receiving a weak
signal with high data
loss from the AP.
Condition exists for
multiple devices in a
specific area
Hard
Transceiver malfunction
Hard
RF Auto Shutoff after 10
minutes of all leads off and
no SpO
What to do
The MMS in use does not
support synchronization of
ECG and SpO2 settings
between the monitor and
Information Center after a
transceiver has been paired.
The MMS must be revision
E.00 or above.
Make sure the transceiver is
the correct revision.
Replace and contact Service.
sensor connected.
2
Technical Alarms (INOPs)
Return patient to the
coverage area.
If patient is in close
proximity to AP,
replace telemetry pack.
Contact service.
The AP covering the
specific area is
suspect. Contact
Service.
Reattach ECG leads to
patient.
Reattach SpO2 sensor.
Press the Check
button.
Alarms
5-19