Philips IntelliVue TRx Instructions For Use Manual page 85

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

In the following table, technical alarms are listed alphabetically.
Alarm Text
BATTERY LOW
BATTERY LOW T
Source - Telemetry
CANNOT ANALYZE
ECG
Source - Information
Center
CANNOT ANALYZE QT
Source - Information
Center
CANNOT ANALYZE ST
Source - Information
Center
Draft - 1 Aug 08
Technical Alarms (Inoperative Conditions)
Priority
Condition
Soft
Battery power is low.
Hard
Arrhythmia algorithm
cannot reliably analyze the
ECG data on any
monitored leads.
Soft
There have been two
consecutive invalid 5
minute values.
Soft
ST algorithm cannot
reliably generate any valid
ST values on any
monitored lead.
Technical Alarms (INOPs)
What to do
There is less than 15 minutes
of monitoring time
remaining. Replace batteries
promptly to avoid transceiver
shutdown and cessation of
monitoring.
Assess the lead selections,
initiate relearn, and validate
analyzed rhythm. Check
other INOPs for possible
source of problem.
See "Limitations for QT
Monitoring" on page 7-23.
Review the ECG signal
quality and correct if
necessary. Reposition the
ISO and J points.
Note—
If the patient has a
ventricular pacemaker, ST
analysis is not possible.
Alarms
5-11