Battery

The battery door must be closed during defibrillation.
Use Duracell Alkaline Batteries, size AA, MN 1500, 1.5V, to ensure
specified performance. Outdated, mismatched, or poor-quality batteries
can give unacceptable performance (e.g., insufficient Battery-Low warning
time). The use of fresh high-quality alkaline batteries is strongly
recommended.
Certain failure conditions, such as short circuits, can cause a battery to
overheat during use. High temperatures can cause burns to the patient and/
or user. If the transceiver becomes hot to the touch, place it aside until it
cools. Then remove the batteries and discard them. Have the transceiver
operation checked by your service provider to identify the cause of
overheating.
If you receive a BATTERY LOW alarm, the batteries must be promptly
replaced. A "Battery Low" condition that is not corrected will result in a
transceiver shutdown and cessation of monitoring.
Draft - 1 Aug 08
Warning
Warning
Warning
Warning
Warning
Warning
Warning
Warning

Battery

Product Safety
2-5