Interference leading to inaccurate measurements can be caused by:
- High levels of ambient light (Hint: cover application site with opaque
material)
- Electromagnetic interference
- Excessive patient movement and vibration.
This equipment is not suitable for use in the presence of a flammable
anesthetic mixture or oxygen concentrations greater than 25% (or partial
pressures greater than 27,5 kPa /206.27 mmHg).
Disposable SpO
become wet. Wet sensors must be replaced immediately.
SpO
The pulse oximeter is calibrated to indicate functional oxygen saturation
2
Information
(fractional oxyhemoglobin), and displayed results can range from 0 to 100%.
for the User
A 10 second averaging filter is used in the calculation of the result. Displayed
results are typically updated every second, but the update period can be
automatically delayed by up to 30 seconds in the presence of noise.