Philips IntelliVue TRx Instructions For Use Manual page 189

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Unit Settings - Telemetry Setup
Control
Function
Suppress
Enable/disable the SpO
SpO
Inops
algorithm to detect NBP running
2
with NBP
and suppress sending technical
alarms from the transceiver for
60 seconds.
Warning
Warning
If you measure SpO
that has an inflated NBP cuff,
a non-pulsatile SpO
alarm can occur. If the
monitor is configured to
suppress this alarm, there can
be a delay of up to 60 seconds
in indicating critical patient
status, such as sudden pulse
loss or hypoxia.
Pleth Wave
Enable/disable the transmission
of the Pleth wave and its
subsequent display to the
Information Center. For
Continuous mode only.
Pulse
Enable/disable the transmission
of the Pulse rate and its
subsequent display to the
Information Center. For
Continuous mode only.
Draft - 1 Aug 08
2
on a limb
2
technical
2
Telemetry Functions at the Information Center & TeleMon
Telemetry Functions at the Information Center
Settings
enable
disable
enable
disable
Note—
When enabled, during
6-lead monitoring, the Pleth
wave will replace the Vb
wave in the Patient Window.
enable
disable
Factory
Default
enable
disable
(Pleth wave
is not
displayed.)
disable
(Pulse rate is
not
displayed.)
9-13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Intellivue trx+

Table of Contents