Philips IntelliVue TRx Instructions For Use Manual page 184

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Telemetry Functions at the Information Center
Patient-Configurable Settings in Telemetry Setup
Control
Function
Suppress
Enable/disable the SpO
SpO
INOPs
algorithm to suppress sending
2
with NBP
technical alarms from the
transceiver during an NBP
measurement for 60 seconds.
Warning
Warning
If you measure SpO
that has an inflated NBP cuff, a
non-pulsatile SpO
alarm can occur. If the monitor
is configured to suppress this
alarm, there can be a delay of
up to 60 seconds in indicating
critical patient status, such as
sudden pulse loss or hypoxia.
Pleth Wave
Enable/disable the transmission
of the Pleth wave (and its
subsequent display) to the
Information Center. For
Continuous SpO
Note—
wave replaces the Vb wave in the
Patient Window during 6-lead
monitoring.
Pulse
Enable/disable display of the
Pulse rate at the Information
Center. For Continuous SpO
mode only.
9-8
Telemetry Functions at the Information Center & TeleMon
Draft - 1 Aug 08
2
on a limb
2
technical
2
mode only.
2
When enabled, the Pleth
2
Setting Choices
enable
disable
enable
disable
enable
disable
Factory
Default
enable
disable
(Pleth is not
displayed.)
disable
(Pulse rate is
not
displayed.)