Draft - 1 Aug 08
Step
Action
3
Test the transceiver:
a. Put fresh batteries in the transceiver (without a leadset
attached) and close the battery door
Result: All six lead lights should flash, and one light
should remain on.
b. Attach a leadset to the ECG port, and attach an SpO
sensor to the SpO
able, attach the ECG leads to the simulator and the SpO
sensor to yourself. At TeleMon, set the SpO
rate to Continuous.
Result: An ECG trace and SpO
visible on the Information Center display. All transceiv-
er lights should be off.
c. Disconnect the Right Arm lead for standard ECG or the
"I" electrode for EASI.
Result: The RA LED or the "I" lead LED should turn
on, and a Leads Off INOP should appear on the display
at the Information Center.
d. Reconnect the electrode.
4
a. Connect the transceiver to TeleMon and observe the
ECG waveform and SpO
play.
Result: The ECG waveform and SpO
be displayed on the TeleMon screen.
port. If an ECG simulator is avail-
2
numerics on the TeleMon dis-
2
Maintenance, Cleaning & Troubleshooting
Cleaning and Sterilization
2
sample
2
information should be
2
numerics should
2
11-13
2