Spot Check Measurement - Philips IntelliVue TRx Instructions For Use Manual

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Spot Check
When the transceiver is configured for Spot Check measurement, use the
Measurement
following instructions to take an individual, manual SpO
transceiver.
Note—
at the Information Center by selecting the Spot Check SpO
"Patient Configurable Settings in Telemetry Setup" on page 9-6).
Note—
connected to TeleMon because only Continuous measurement is available with
this device.
Draft - 1 Aug 08
You can also initiate a Spot Check measurement in the Patient Window
Spot Check measurements cannot be made when the transceiver is
Step
Action
1
Attach the sensor to the patient.
2
Connect the SpO
The SpO
A low-pitch tone detecting each pulse is audible (unless
sounds are muted).
3
After approximately 30 seconds, if the measurement was successful,
you'll hear a single beep from the transceiver. The value, with the
measurement time, will be displayed at the Information Center. The
sensor light will extinguish.
If the measurement was unsuccessful, you'll hear a double beep.
Remove the sensor cable and reinsert it to retake the measurement.
The SpO
Note—
Center for 1 hour or until another measurement is taken, with one
exception: If the batteries are removed from the transceiver, the Spot
Check measurement will be erased from the display; however, the
SpO
measurements will be available in Trend Review.
2
4
To repeat a Spot Check measurement at the bedside, disconnect
then reconnect the SpO
cable to IntelliVue TRx
2
sensor light turns on.
2
value and time stamp remain on the Information
2
cable to the transceiver.
2
Measuring SpO
reading from the
2
icon (see
2
+
, and check that:
SpO
Monitoring
8-17
2
2