Testing Transceiver Functionality
Status
You can check the status of the transceiver indicators at any time.
Check
Initiating a

Status Check

4-12
Basic Operation
Draft - 1 Aug 08
Step
Action
1
Press the
Check
The following indicators should illuminate for as long as the Check
button is depressed.
Battery gauge
Leads Off indicators
EASI (if in use)
2
If one or more of the expected indicators do not light up, check the
following:
Power and position of batteries (see "Checking the Battery
Power Level" on page 4-17).
Leadset insertion. Make sure the leadset is correctly inserted
in the transceiver and the colored line at the base of the cable
is not visible (see "Connecting the ECG Cable" on page 6-
21).
Lead positions and connections (see "Verifying Electrode
Connections" on page 6-25).
If there is still a problem, contact your service provider for
assistance.