Philips IntelliVue TRx Instructions For Use Manual page 149

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Draft - 1 Aug 08
Adjustment
Set the QT Baseline
How the QT Analysis Algorithm Works
Description
To quantify changes in the QTc value you can
set a QTc baseline. Select the Set QT Baseline
button to replace the baseline QTc value with the
current QTc value. If a baseline has not been set
the Information Center sets the baseline to the
first valid value after measuring QT for five
minutes. Setting a new baseline discards the
previous baseline.
Notes—
• Since the dQTc alarm is based on the
difference between the baseline and the
current value, setting an inappropriate new
baseline may prevent a dQTc alarm from
being generated.
• Turning QT analysis off does not clear the
baseline value. This allows you to turn QT
analysis off during prolonged arrhythmias,
such as bigeminy, without losing the
baseline.
• Discharging a patient clears the baseline.
ST/AR Arrhythmia Monitoring
7-21

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Intellivue trx+

Table of Contents