Philips IntelliVue TRx Instructions For Use Manual page 219

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Condition
Replace BatteryT INOP
displayed at the monitor
and the Information Center.
Absence of successful
association sound/visual
indicator.
Central: Tele Only INOP
Check Pairing INOP
Draft - 1 Aug 08
Description
When operating wirelessly, the
bedside monitor is no longer
providing power to the
transmitter, and battery capacity
is now depleted. There is no
monitoring occurring at either
the monitor or the Information
Center.
The bedside monitor and the
transceiver are not
communicating with each other.
Monitoring at the Information
Center only.
When the transceiver is wirelessly
paired with an X2 patient monitor
(no label) docked with a larger
networked MP series monitor,
and the network connection is
lost.
An X2 patient monitor with a
telemetry label is paired with a
larger networked MP series
monitor and subsequently the
transceiver is paired with the
same monitor. Only one
transceiver can pair with the
monitor.
Pairing Non-networked Devices
Action
Insert new AA batteries in
the transceiver.
Repeat the device
assignment procedure.
Restore the monitor's
network connection.
Select the correct device to
be paired.
Pairing Monitoring Devices
10-23