6.2.9.3.

SAV_triggering pressure

For the measurement, it is important to ensure that the valve
upstream of the SSV is open and the valve upstream of the
consumer is closed. The pressure downstream of the controller
(secondary side) in the gas pipe must be increased. If the triggering
pressure is exceeded, the SSV is triggered, indicated by a click,
and the gas supply is blocked on the primary side. The existing
pressure at the time of the click (secondary side) is the SSV
triggering pressure.
Calling up the function:
>
[
1.
SAV_triggering pressure
-
Measurement is started
-
Gas is released from the pipe into the gas feeding unit. Carry
out steps in accordance with the information on the display and
confirm with [OK].
-
Zeroing starts (5 s).
-
The pressure is increased automatically by the instrument.
The following values are displayed:
Parameter
Pressure
Max.
2. End the measurement as soon as the safety shut-off valve
closes:
3. Save measurement result: [Save].
Options
>
[Options]
values is displayed in a line graph (zoom function available [▲],
[▼]).
>
[Options]
in the same way.
>
[Options]
across the overall measuring time are displayed in a line graph.
The SSV stop must then be manually released again.
]
Measurements
Explanation
Current pressure
Highest pressure during the current
measurement
[
].
Show
Graphic: a section of past measuring
→ Manometer: the measuring values are displayed
Overall graphic
[OK]
Controller test
→ [OK].