Differential Pressure; Differential Temperature - TESTO 300 Pro Smoke Edition Manual

Combustion analyzer
Table of Contents

Advertisement

Find Quality Products Online at:

9.6.6 Differential pressure

Dangerous mixture of gases!
Danger of explosion!
- Make sure there are no leaks between the sampling point and the
measuring instrument.
- Do not smoke or use naked flames during the measurement.
The gas pressure kit (0554 1203) must be connected.
For an instrument with no dilution option: The minus connection for
differential pressure measurement must be depressurized at the start of
the measurement (ambient pressure, instrument not connected to
system being checked), since the pressure sensor is zeroed.
1
Call up function:
|
2
Tap
Differential
pressure.
3
Start measurement: tap
Zeroing the pressure sensor.
Reading is displayed.
4
End measurement: tap

9.6.7 Differential temperature

The differential temperature kit (0554 1208) or two external temperature
probes must be connected.
1
Call up function:
|
2
Start measurement: tap
The readings and the calculated differential temperature
are displayed.
3
End measurement: tap
GlobalTestSupply
www.
9 Performing the measurement
WARNING
Differential pressure
.
.
Differential temperature
.
.
ΔT