Carrying Out Repeated Testing Ps <0.5 Bar - TESTO 324 Instruction Manual

Leakage rates measuring instrument
Table of Contents

Advertisement

6.2.5.4.
Carrying out repeated testing PS <0.5 bar
The aim of this testing is to establish, at the time of testing, whether
the pipe is in proper working condition for the intended application
and can satisfy the requirements until the next test.
Call up a function:
>
[
Repeated testing PS < 0.5
Setting parameters
Parameter
Stability time
Measurement period
Test pressure (target) Set test pressure
1. Edit values: [Edit].
2. Enter values. [▲],
Carrying out repeated testing (PS <0.5 bar)
1. Start measurement:
-
The test pressure (target) is built up automatically by the
instrument.
-
Zeroing starts (5 s).
Option
> End stability time early: [Next].
> End measurement period early: [Next].
-
-
Pressure test is ended.
2. Evaluate measurement result: [▲],
-
The measurement result is displayed.
3. Save measurement result: [Save].
]
Measurement options
Explanation
Set stability time
Set measurement period
[▼]
[
Measurement is cancelled.
[OK]
bar.
DVGW TRF 2012 and
ÖVGW G2/2 specify a
measurement period of 10 min.
DVGW TRF 2012 and
ÖVGW G2/2 specify a test
pressure of 150 mbar.
and sometimes [◄],
].
[▼]
→ [OK].
6 Using the product
Liquid gas (TRF)
[►]
→ [OK].
55