6.6.2 Total Chlorine Measurement Programming

6.6.2.1 Description
This Chlorine sub-section describes how to configure the Model 1056 analyzer for Total Chlorine measurement using
amperometric chlorine sensors. The following programming and configuration functions are covered:
TABLE 6-6. Total Chlorine Measurement Programming
Measure
Sec.
Total
6.6.2.2
Chlorine
6.6.2.3
6.6.2.4
6.6.2.5
A detailed flow diagram for programming of all chlorine measurements is provided at the end of Sec. 6 to
guide you through all basic programming and configuration functions.
To configure the chlorine measurement board for total
chlorine:
1.
Press MENU
2.
Scroll down to Program. Press ENTER.
3.
Scroll down to Measurement. Press ENTER.
4.
Select Sensor 1 or Sensor 2 corresponding to
chlorine. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any displayed function, scroll
to the desired item and press ENTER.
The following sub-sections provide you with the initial display screen that appears for each configuration function.
Use the flow diagram for chlorine programming at the end of Sec. 6 and the Model 1056 live screen prompts for
each function to complete configuration and programming.
6.6.2.2 Measure
The display screen for selecting the measurement is
shown. The default value is displayed in bold type.
Refer to the chlorine Programming flow diagram to
complete this function.
6.6.2.3 Units
The display screen for selecting units as ppm or mg/L
is shown. The default value is displayed in bold type.
Refer to the Chlorine Programming flow diagram to
complete this function
Menu function:
default
Measure:
Free Chlorine
Units:
ppm
Filter:
5sec
Resolution:
0.001
Section 6.0: Programming the Measurements
Description
Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine
Select units ppm or mg/L
Override the default input filter, enter 0-999 seconds
Select 0.01 or 0.001 display resolution
S1: 1.234µS/cm
S2: 12.34pH
S1: 1.234µS/cm
S2: 12.34pH
S1: 1.234µS/cm
S2: 12.34pH
25.0ºC
25.0ºC
SN Configure
Measure: Free Chlorine
Units:
ppm
Filter:
5sec
Resolution:
0.001
25.0ºC
25.0ºC
SN Measurement
Free Chlorine
pH Independ. Free Cl
Total Chlorine
Monochloramine
25.0ºC
25.0ºC
SN Units
ppm
mg/L
April 2017
55

Advertisement

Table of Contents
loading

Table of Contents