6.7 Oxygen Measurement Programming

6.7.1 Description
This section describes how to configure the Model 1056 analyzer for dissolved and gaseous oxygen measurement
using amperometric oxygen sensors. The following programming and configuration functions are covered:
TABLE 6-9. Oxygen Measurement Programming
Measure
Sec.
Oxygen
6.7.2
6.7.3
6.7.4
6.7.5
6.7.6
6.7.7
6.7.8
A detailed flow diagram for oxygen programming is provided at the end of Sec. 6 to guide you through all basic pro-
gramming and configuration functions.
To configure the Oxygen measurement board:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to Oxygen.
Press ENTER.
The adjacent screen format will appear (factory defaults are
shown). To program any displayed function, scroll to the
desired item and press ENTER.
The following sub-sections show the initial display screen that appears for each configuration function. Use the
flow diagram for oxygen programming at the end of Sec. 6 and the Model 1056 live screen prompts for each
function to complete configuration and programming.

6.7.2 Oxygen Measurement application

The display screen for programming the measurement
is shown. The default value is displayed in bold type.
Refer to the Oxygen Programming flow diagram to
complete this function.
6.7.3 Units
The display screen for selecting units as ppm , mg/L,
ppb, µg/L, % Saturation, %Oxygen in Gas, or ppm
Oxygen in Gas is shown. The default value is displayed
in bold type. Refer to the Oxygen Programming flow
diagram to complete this function.
Menu function:
default
Type:
Water/Waste
ppm Select ppm, mg/L, ppb, µg/L, % Sat, %O2-Gas, ppm Oxygen-Gas
Units:
Partial Press:
mmHg
Salinity:
00.0‰
Filter:
5sec
Pressure Units:
bar
Use Press:
At Air Cal
Section 6.0: Programming the Measurements
Description
Select Water/Waste, Trace. BioRx, BioRx-Other, Brew, %O2 In Gas
Select mm Hg, in Hg. atm, kPa, mbar or bar for Partial pressure
Enter Salinity as ‰
Override the default input filter, enter 0-999 seconds
Select pressure units: mm Hg, in Hg,. Atm, kPa, mbar, bar
Select atmospheric pressure source – internal or mA Input
S1: 1.234µS/cm
S2: 12.34pH
Type:
Units:
Partial Press: mmHg
Salinity:
Filter:
Pressure Units:
Use Press: At Air Cal
Custom Setup
S1: 1.234µS/cm
S2: 12.34pH
Water/Waste
Trace Oxygen
BioRx-Rosemount
BioRx-Other
Brewing
Oxygen In Gas
S1: 1.234µS/cm
S2: 12.34pH
ppm
mg/L
ppb
µg/L
% Saturation
Partial Pressure
% Oxygen In Gas
ppm Oxygen In Gas
25.0ºC
25.0ºC
SN Configure
Water/Waste
ppm
00.0‰
5sec
bar
25.0ºC
25.0ºC
SN Type
25.0ºC
25.0ºC
SN Units
April 2017
59