(4) Checking the Sensor Serial No.
1)
Purpose
If the sensor serial number and the sensor serial number stored in the flash ROM of the
50Di PCB differ from the image data file name stored in the hard drive of the Control PC
due to replacing the PCB or imaging unit, the connected sensor can not be detected after the
CXDI application is launched. In that case an error message appears.
2)
Notes
2-1) Check the sensor serial No. whenever:
a. 50Di PCB is replaced.
b. Sensor in the imaging unit is replaced.
2-2) This checking procedure must be performed with the Control PC, Imaging Unit and
all the other equipments connected and started up.
3)
Procedure
3-1)
Check the sensor serial No. by referring to "Sensor serial number setting" in the
Tool Software Operation Manual for CXDI-50G, and write the number as required.
2. Installation Manual
- 33 -