Canon CXDI-50G Service Manual page 163

Cxdi-1 system
Hide thumbs Also See for CXDI-50G:
Table of Contents

Advertisement

Error Code
Reply
Seriousness
Keys
-126
OK
Error
-129
OK
Error
0xEFF0
0xEFEF
-143
OK
Error
Error Code List for Version 6.00
[Operation after Reply Key is touched]
#1:Error
#2:Sensor Unit:Temperature Error
#3:The internal temperature of the Sensor Unit exceeded the
temperature limit.
Cannot capture images until the temperature goes below
the limit.
#4:TEMP OVER (n)(effa)
n: Sensor Index (1 to 4)
[Cause]
The sensor temperature has exceeded 45°C. Thus, the error
for aborting the operation has been detected. The system
cannot be used until the temperature lowers, which will take
several tens of minutes.
[Operation after Reply Key is touched]
The sensors automatically enter the sleep mode. The host
computer does nothing. If you try to make the system ready,
the same error occurs again.
Press the [OK] to close the window, after few minutes later
operate again.
Note: The error might be able to be evaded by reducing an
environmental temperature.
#1:Error
#2:Sensor Unit: Data Transmit Error
#3:Failed to transmit the data.
[Cause]
Internal DMA has passed time limit during read the reduced
image.
Full image data transferring has been interrupted.
[Operation after Reply Key is touched]
Without anything to do. (Same as usual operation)
#1:Error
#2:Calibration Table Error
#3:Perform Calibration.
[Cause]
No gain calibration table has been prepared. Or, the serial
number of the connected sensor unit is different.
[Operation after Reply Key is touched]
Perform calibration to make a gain calibration table.
- 8 -
ror Code List for Versi
Message
[Cause]