Error Code
Reply
Seriousness
Keys
-123
OK
Error
Cancel
0xeff8
0xeff7
-125
OK
Error
0xeffb
Error Code List for Version 6.00
[Operation after Reply Key is touched]
#1:Error
#2: Sensor Unit: PLD Error
#3: Detected PLD Error. Call a service person.
[Cause]
A PLD error has been found in the sensor A/D board.
[Operation after Reply Key is touched]
When OK key is touched, the sensor will be disconnected.
(Access will not be made, as the system is in an "abnormal"
state.)
If this error persists even after the system is rebooted, a
service person needs to be called in.
"PLD err (n)( eff8)"
"PLD Config err (n)( eff7)"
n: Sensor Index (1 to 4)
#1:Error
#2:Sensor Unit:Grid Error
#3:Because of the Grid error, this image may have a quality
problem.
[Cause]
Although a warning for grid had been detected in the sensor
A/D board, imaging was continued. Thus, the image may
have been degraded.
[Operation after Reply Key is touched]
When OK key is touched, the screen returns to the normal
display without doing anything.
"Grid was changed during exposure (n)(effb)"
n: Sensor Index (1 to 4)