Available Imaging Time (Normal Imaging/Long-Term Imaging); Overview; Precaution; Technical Description - Canon CXDI-50G Service Manual

Cxdi-1 system
Hide thumbs Also See for CXDI-50G:
Table of Contents

Advertisement

A
v
a
i
l
a
b
l
e
I
m
a
g
A
v
a
i
l
a
b
l
e
I
m
a
g
3
3

[Overview]

There are two available imaging times as follow:
1) Normal imaging (0 - 1000 msec)
2) Long-term imaging (1001 - 3000 msec)
From Version 6.0, the maximum imaging time is now 3 seconds for the CXDI-50G sensor,
which can be set on the GUI.

[Precaution]

• The preview display becomes slower when using long-term imaging.
• The preview display may also become slower
part button that is set to long-term imaging (imaging time sett ings: 1001 msec to 3000 msec).
*1
The preview display becomes slower without the X-ray monitor, that is, when using the
CXDI-50G.
(When an X-ray exposure negative cannot be detected, even if the X-ray monitor is used.)
• Imaging preparation time: 10 seconds for normal imaging and 30 seconds for long-term
imaging. In imaging preparation time, the imaging unit changes from sleep status (sensor in
standby status) to ready status (possible to perform imaging).
(The imaging cycle time
*2
Time required preparing for the next imaging after a single imaging is done. Normal
imaging: 15 sec, long-term imaging: 30 sec.

[Technical Description]

1. The maximum value for imaging condition
• When using the CXDI-50G sensor, the maximum value of the exposure time is 3000
msec, which can be set on the screen.
• The maximum value of the tube current is 1600 mA. - The maximum mAs value is
4800mAs due to the maximum value of the exposure time changing to 3000.
The maximum value of the tube current is 1600mA. The maximum value of the mAs
value is changed from the 3000 maximum value of the exposure time to 4800 mAs.
Up to CXDI V5.1
CXDI-11- CXDI-31 sensor
CXDI-40G or later version sensor
2. Imaging condition check at the time of upgrade
If any of the kV, mA, mAs, cm (except msec) values recorded in the imaging condition
table or in each body part information exceeds the limit of the sensor, when the CXDI
starts up, a warning message will be displayed (Fig. 1) and the value will be automatically
modified to be at or below the limit. (However, thinking from the range of values that can
3. Function
i
n
g
T
i
m
e
(
N
o
r
m
a
i
n
g
T
i
m
e
(
N
o
r
m
a
*2
becomes longer.)
Exposure time
1000msec
1000msec
3000msec
Table 3-1
l
I
m
a
g
i
n
g
/
L
o
n
g
-
t
e
l
I
m
a
g
i
n
g
/
L
o
n
g
-
t
e
*1
when performing normal imaging using the
Tube current
5000mA
1600mA
1600mA
5
r
m
I
m
a
g
i
n
g
)
r
m
I
m
a
g
i
n
g
)
mAs value
5000mAs
1600mAs
4800mAs