Error Code
Reply
Seriousness
Keys
-6
OK
Alert
-8
OK
Alert
-116
OK
Alert
-5002
OK
Alert
Error Code List for Version 6.00
[Operation after Reply Key is touched]
#1:Alert
#2:System Info Error
#3:A/D board info is updated.
[Cause]
The A/D board information is different from the former
information. This problem occurs when the version of the
A/D board has been updated or when the connected units have
been changed. If this error occurs when no such changes have
been made, it is attributed to another problem. In this case, a
service person needs to be called in.
[Operation after Reply Key is touched]
The A/D board information is renewed.
#1:Alert
#2:System SelfCheck Error
#3:Call a service person and let him/her know the error code.
[Cause]
An error has been found by a self-check.
[Operation after Reply Key is touched]
Although imaging is possible, image may be degraded.
Therefore, a service person needs to be called in.
#1:Alert
#2:Sensor Unit:SelfCheck Error
#3:Detected an Error.
#4:ER(n) (xxxx)
n: Sensor Index (1 to 4)
xxxx: Error code
[Cause]
An undefined error has been found in the sensor A/D board.
[Operation after Reply Key is touched]
Nothing
#1:Alert
#2:OPU-CCR RPC Error
#3:OPU-CCR RPC Error.
[Cause]
For some reason, the CCR software has shut down.
[Operation after Reply Key is touched]
- 4 -