Canon CXDI-50G Service Manual page 170

Cxdi-1 system
Hide thumbs Also See for CXDI-50G:
Table of Contents

Advertisement

Error Code
Reply
Seriousness
Keys
-5101
OK
Error
-5102
OK
Error
Error Code List for Version 6.00
[Operation after Reply Key is touched]
#1:Error
#2:Sensor Unit:Detect Error
#3:Check the sensor unit cable connection. If you want to
recover the system, shut it down.
[Cause]
The cable is not connected though it is setting which uses the
sensor.
[Operation after Reply Key is touched]
The system only requires your confirmation. Touch OK. Only
the problematic sensor is disabled.
#1:Error
#2:Grid Error
#3:Remove Grid for this exposure.
[Cause]
This error occurs when grids are detected in the
Buck/Cassette setting, where imaging is expected to be
performed without grids.
[Operation after Reply Key is touched]
Touch OK after the check.
- 15 -
ror Code List for Versi
Message
[Cause]

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents