Canon CXDI-50G Service Manual page 172

Cxdi-1 system
Hide thumbs Also See for CXDI-50G:
Table of Contents

Advertisement

Error Code
Reply
Seriousness
Keys
-153
OK
Warning
-502
RETRY
Warning
CANCEL
-503
RETRY
Warning
CANCEL
Error Code List for Version 6.00
[Operation after Reply Key is touched]
#1:Warning
#2:Sensor Unit:Temperature Warning
#3:The internal temperature of the Sensor Unit approaches the
temperature limit.
Do not leave in the READY condition of the Sensor Unit..
#4:TEMP OVER (n)(eff9)
n: Sensor Index (1 to 4)
[Cause]
The warning sign is detected because the internal temparature
of Imaging unit
has approached the upper bound value.
[Operation after Reply Key is touched]
The sensor selfishly enters the state of sleep. There is nothing
to operate at the host side.
The same error occurs when the user tries to put this sensor
into the state of READY.
In this case, press the [OK] to close the window and continue
to operate. This error might be able to be evated by reducing
the internal temparature or having the Imaging unit READY
STATE TIME shorten.
#1:Warning
#2:DICOM Connect Error
#3:Cannot connect to target. Check network or port number
setting.
[Cause]
Cannot be connected to the destination.
[Operation after Reply Key is touched]
When RETRY is chosen, the image will be transferred again.
When CANCEL is chosen, transfer will be canceled after
displaying a confirmation message. Re-transfer is possible by
sending the image from the study list.
#1:Warning
#2:DICOM Transfer Error
#3:Error has occurred during association.
[Cause]
Network was cut off during the communication.
[Operation after Reply Key is touched]
When RETRY is chosen, the image will be transferred again.
When CANCEL is chosen, transfer will be cancelled after
displaying a confirmation message. Re-transfer is possible by
transferring the image from the study list.[990914]