4. Repair Guide
Removing the PCB-50 LED
1.5
Screw x2
Grid detect sensor
PCB-50 LED
Connector
Screw x4
1. After removing the connector connected with the PCB-50LED unit, loosen the screw (x2)
fixes the Grid detect sensor on the top cover and remove this screw.
2. Loosen the screw (x4) fixes the PCB-50 LED on the top cover and remove this PCB.
Note:
1. At the assembling, draw the cable carefully so as not to crowd.
2. Do not tighten the screw so much.