Error Code
Reply
Seriousness
Keys
-144
OK
Error
-302
OK
Error
-303
OK
Error
-304
OK
Error
Error Code List for Version 6.00
[Operation after Reply Key is touched]
#1:Error
#2:Sensor Unit Configuration
#3:Incorrect Senor Unit Configuration.
Call a service person.
#4:Invalid Serial No.(n)(eff5 or eff6)
n: Sensor Index (1 to 4)
[Cause]
0xeff5: Caused by that the differences has been in serial
number and DP number.
0xeff6: Caused by that false serial number has written.
[Operation after Reply Key is touched]
When OK button is touched, sensor is disconnected. A service
person needs to be called in. (Keep unusual state, and do not
access.)
#1:Error
#2:Calibration Error
#3:Too Much Dose.
Or collimator is used.
[Cause]
Excessive dose of radiation was used in exposure of image for
white correction.
[Operation after Reply Key is touched]
Perform imaging again.
#1:Error
#2:Calibration Error
#3:Too Little Dose.
Or collimator is used.
[Cause]
Insufficient dose of radiation was used in exposure of image
for white correction.
[Operation after Reply Key is touched]
Perform imaging again.
#1:Error
#2:Calibration Error
#3:Invalid Data. Or collimator is used.
[Cause]
Too much difference in dose acquired by the panels.
[Operation after Reply Key is touched]
Perform imaging again.