Canon CXDI-50G Service Manual page 11

Cxdi-1 system
Hide thumbs Also See for CXDI-50G:
Table of Contents

Advertisement

Item
Imaging restriction
(Imaging Prohibition)
User interface
1. General
Content
When the internal temperature of
Imaging Unit is above 49 degree
Celsius, its state is changed to sleep
mode. And the Imaging prohibition
will be continued when the internal
temperature is below 48 degree
Celsius.
Single type LED
Off: Imaging unit power is off
On in orange: Imaging unit power is
on (unable to perform imaging)
Blinking in green: Preparing
imaging/error status
On in green: Imaging preparation is
complete
INDICATIO
N
Color
Imaging unit
is off
Imaging unit
is on
Preparing
imaging
Imaging
preparation
complete
Error status
Communi-
cating
Initialization
(when
startup)
Network not
set (when
startup)
*1: Turns on and off for 0.5 seconds each
*2: Turns on and off twice for 0.5 seconds, then
turns off for 0.5 seconds
*3: Turns on and off randomly
*4: Fades in for 1 second and fades out for 1
second
*5: Fades in for 2 seconds, and then turns off
5
Remarks
BUSY
SENSOR
POWER
Orange
Green
Off
Off
--
Off
*1
--
Blinking
--
Off
*2
--
Blinking
*3
Blinking
--
*4
--
Blinking
*5
--
Blinking
Blue
Off
On
On
On
On
On
On
On