www.norrscope.com
Technical data
9.5
Physical specifications
Enclosure
copper free aluminum or 316 stainless steel
material
Enclosure
NEMA 4X and IP 65/66/67
protection
Display
4 digit, 9 segment, back-lit
Size (LxWxD,approx.)
Without Docking Station
11.3" x 5.8" x 5.1"
With Docking Station
11.5" x 7.1" x 7.4"
Weight (approx.)
Without Docking Station
aluminum
stainless steel 316
With Docking Station
aluminum
stainless steel 316
9.6
Environmental parameters
Pressure
20.7 to 38.4 in. of Hg (700 to 1300 hPa)
Humidity
0% to 100% RH, non-condensing
Temperature -40 to +149 °F (-40 to +65 °C)
Storage
-4 to +149 °F (-20 to +65 °C)
temperature
See sensor data sheet for sensor specifications
9.7
Environmental influences
For influences on the measurement performance and
restrictions of a particular sensor see sensor data sheet.
28
9.8
The values are valid for aluminum and stainless steel 316
versions.
(285 x 150 x 130 mm)
(295 x 180 x 190 mm)
Grounding screws
9.9
4.9 lbs. (2.2 kg)
8.8 lbs. (3.6 kg)
7.7 lbs. (3.5 kg)
11.9 lbs (5.4 kg)
supply and
Tightening torque for instrument threads
Part
TQ Lb. In.
Enclosure lid
min. 44.3
Sensor
min. 266
Blind plug
min. 266
Feed-through
min. 443
Conduit hub
Instrument to
Docking Station
71
(4 screws)
10.6
Tightening torque and size for field
wiring terminals
Terminal
TQ Lb. In.
Power
4.4 – 7.0
(0.5 – 0.8 Nm)
signal
4.4 – 7.0
Relays
(0.5 – 0.8 Nm)
TQ Nm
min. 5
min. 30
min. 30
min. 50
8
1.2
Wire Size
Wire Size
2
AWG
mm
24 – 12
0.2 – 2.5
20 – 12
0.5 – 2.5