Ssb Phase Noise Sensitivity Test (For Standard); Ssb Phase Noise Sensitivity Test (For Option 011); Sa Relative Level Accuracy Test - Agilent Technologies E5052A Service Manual

Signal source
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

Performance Tests
E5052A Performance Test Record
Frequency [Hz]
Test limit [dB]
2.0125 G
3.0125 G
4.0125 G
5.0125 G
6.0125 G
7 G

SSB Phase Noise Sensitivity Test (for Standard)

CW Frequency: 70 MHz, Correlation = 1, Start frequency = 1 Hz
Offset Frequency
Test limit [dB/Hz]
[Hz]
1 M
3 M
7 M

SSB Phase Noise Sensitivity Test (for Option 011)

CW Frequency: 70 MHz, Start frequency = 10 Hz
Offset Frequency
Test limit [dB/Hz]
[Hz]
1 M
3 M
7 M

SA Relative Level Accuracy Test

CW Frequency: 100 MHz
Offset Frequency
[Hz]
-7.5 M
-6.5 M
-5.5 M
70
Test result
± 0.5
± 1.0
± 1.0
± 1.0
± 1.0
± 1.0
Test result
< -168.5
< -169.0
< -169.7
Test result
< -160.0
< -160.0
< -160.0
Test limit [dB]
Test result [dB]
± 1.5
± 1.5
± 1.5
[dB]
Measurement
uncertainty [dB]
± 0.13
± 0.13
± 0.16
± 0.16
± 0.17
± 0.19
[dB/Hz]
[dB/Hz]
Measurement
uncertainty [dB]
± 0.17
± 0.17
± 0.16
Chapter 2