To Check The Device Driver - Agilent Technologies E5052A Service Manual

Signal source
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

Table 4-2
Major functions and troubleshooting procedures
Function
Printer parallel port
GPIB Interface
24-bit I/O port
Ref Oven
USB (USBTMC)
Interface

To Check the Device Driver

Make sure first whether the E5052A device drivers are installed properly or not by the
following procedure, if a function of specific device in the E5052A doesn't work.
Step 1. Exit from the E5052A measurement view in accordance with the procedure described in
"How to exit from the E5052A Measurement View" on page 85. Then, Windows desktop
screen is displayed.
Step 2. Click "My Computer" with the right button and select "Properties" as shown in Figure
4-17. Then, the System Properties(Figure 4-18) will appear.
Figure 4-17
Opening System Property Window
Chapter 4
Description
A specified printer can be used to print the
E5052A measurement display, setup display and
others, when it is connected to the Printer port
(36-pin D-Sub) on the rear panel.
The GPIB compatibility allows the E5052A to be
operated as a talker/listener on IEEE 488 interface
bus.
The 24-bit I/O port can be used to transfer a
comparator decision output data to and perform
timing synchronization with an external handler.
A 10 MHz high stability signal is outputted.
The GPIB compatibility allows the E5052A to be
operated as a talker/listener on IEEE 488 interface
bus.
Troubleshooting
Function Specific Troubleshooting
Troubleshooting
Refer to "To Check the
Printer Parallel/USB Port"
on page 110.
Refer to "To Check the
GPIB" on page 111.
Refer to in "To Check the
24-bit I/O port" on page 111.
Refer to in "To Check the
Ref Oven" on page 111.
Refer to in "To Check the
USB (USBTMC) Interface
Card" on page 111.
101