A12 Power Distribution Board Replacement; Tools Required; Removal Procedure - Agilent Technologies E5052A Service Manual

Signal source
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

A12 Power Distribution Board Replacement

Tools Required

TORX screwdriver, T10, T15, and T20

Removal Procedure

Refer to Figure 6-20 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 176.
Step 2. Remove the Power supply as described in "A50 Power Supply Assembly Replacement" on
page 190.
Step 3. Remove the rear plate (item 1).
Step 4. Remove the two TORX T10 screws (item 2) fastening the plate.
Step 5. Disconnect the cables from the Power Distribution board.
Step 6. Remove the four TORX T10 screws (item 3) fastening the Power Distribution board.
Step 7. Slide out the Power Distribution board (item 4).
Figure 6-20
A12 Power Distribution Board Removal
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
Chapter 6
Replacement Procedure
A12 Power Distribution Board Replacement
203

Hide quick links:

Advertisement

Table of Contents
loading