Performance Tests
Individual Test Description

2. Frequency Measurement Test

Description
This test checks the frequency measurement accuracy of the E5052A. The frequency
measurement is checked for each frequency band and each frequency resolution of the
E5052A
Specification
@23 ± 5 °C
Frequency accuracy
Test equipment
Signal Generator
RF Amplifier (AMP 1)
RF Amplifier (AMP 2)
Power Meter
Power Sensor
Step Attenuator
Power Splitter
Blocking Capacitor
N(f)-SMA(m) adapter, 2 ea.
3 dB Fixed Attenuator, 2 ea.
N(m)-N(m) cable 61cm, 3 ea.
N(m)-N(m) Adapter
BNC cable, 122cm
Procedure for Standard
Step 1. Connect the power sensor to the power meter. Perform the zero adjustment and the
calibration for the power meter.
Step 2. Connect the test equipment as shown in Figure 2-2
28
± (resolution + time base accuracy) @ 10 MHz to 7 GHz
Frequency Resolution: 10 Hz, 1 kHz, and 64 kHz
Internal time base stability: ±5 ppm
Agilent E8257D
R&K A3000-2H-R
Agilent 83020A
Agilent E4419B
Agilent 8481A
Agilent 8496H
Agilent 11667A
Agilent 11742A
Agilent p/n 1250-1562
Agilent 8491A Opt.003
Agilent N6314A (p/n 8120-8862)
Agilent p/n 1250-1475
Agilent p/n 8120-1840