Rotors for the accuSpin™ 3
Table 1:
Rotor designation
With bucket
Order no.
Maximum permissible load [ g ]
Maximum speed n
[ rpm ]
max
Maximum RCF value at n
Radius (max.)
[ cm ]
Acceleration / deceleration time [ s ]
Heating of samples at n
relative to room temperature 23°C,
run time 1 hour
Aerosol-tight *
Autoclavable
* Checked by TÜV (Technical Supervision Association) Product Service GmbH - Hamburg
Round bucket
750 ml
7500 4394
4 x 800
4150
3 696
max
19.2
60 / 30
7°C
max
yes
(with cap 75004397)
121°C
(differences of 230V instruments are shown in parentheses)
Swinging Bucket Rotor 7500 4393
Rectangular bucket
250 ml
7500 4395
4 x 750
( 4350 )
3750
( 4083 )
2 986
19.0
( 40 / 40 )
60 / 30
( 9°C )
7°C
yes
(with cap 75006448)
121°C
Rotors and accessories
Carrier for
microplates
7500 4396
4 x 500
( 4350 )
4150
( 4020 )
3061
15.9
( 45 / 40 )
50 / 25
( 10°C )
4°C
no
121°C
( 4350 )
( 3364 )
( 40 / 35 )
( 5°C )
17