VIII. SET THE SAMPLING TIME FOR DATA LOGGING
Press the SETUP button several times until SEC is displayed in reverse video.
Press the ▲ or ▼ buttons to increment or decrement the value (users can set up
the sampling time from 2 sec. to 3000 sec., step by 2, e.g. 2 sec., 4 sec., 6
Holding the ▲ or ▼ buttons will speed up the process of
sec., ... ).
incrementing or decrementing.
To exit, press the EXIT button.
The example about how to decide the sampling time for data logging:
If users want to log data for 3 weeks and the power system is 3P4W, first clear the
memory of the analyzer, then set up the sampling time as 104 sec. or more. See
the calculation on below:
3 (weeks) x 7 (days) x 24 (hours) x 60 (min.) x 60 (sec.) = 1,814,400 (sec.)
1,814,400 (sec.) / 17,474 (the max. file capacity of 3P4W) = 103.83 (sec.)
46