Specifications (23℃± 5℃) - PCE Instruments PCE-830 User Manual

Table of Contents

Advertisement

XI. SPECIFICATIONS (23℃± 5℃)
AC Watt
(50 or 60 Hz, PF 0.5 to 1, CT = 1, Voltage > AC 20V, Current > AC 40mA for 1A
range, Current > AC 0.4A for 10A range, Current > AC 4A for 100A range, and
continuous waveform)
PCE-830 + PCE-6801 (100A)
Range (0 to 100A)
5.0 – 999.9 W
1.000 – 9.999 KW
10.00 – 99.99 KW
100.0 – 999.9 KW
1000 – 9999 KW
(50 or 60 Hz, PF 0.5 to 1, CT = 1, Voltage > AC 20V, Current > AC 0.4A for 10A
range, Current > AC 4A for 100A range, Current > AC 40A for 1000A range, and
continuous waveform)
PCE-830 + PCE-6802 (1000A)
Range (0 to 1000A)
5.0 – 999.9 W
1.000 – 9.999 KW
10.00 – 99.99 KW
100.0 – 999.9 KW
1000 – 9999 KW
0.000 – 9.999MW
(50 or 60 Hz, PF 0.5 to 1, CT = 1, Voltage > AC 20V, Current > AC 12A for 300A
range of 3007, Current > AC 120A for 3000A range of 3007, and continuous
waveform. Conductor is located at the center of flexible loop. Position sensitivity is
2% of range. External field effect of < 40A/m and 200mm from the coupling is 1%
of range. Temperature coefficient is 0.02% of reading /° C)
PCE-830 + PCE-3007 (3000A/1200A)
Range (0 to
3000A/1200A)
10.0 – 999.9 W
1.000 – 9.999 KW
10.00 – 99.99 KW
100.0 – 999.9 KW
Resolution
0.1W
0.001 KW
0.01 KW
0.1 KW
1 KW
Resolution
0.1W
0.001 KW
0.01 KW
0.1 KW
1 KW
0.001MW
Resolution
> 20 V and > 30A
0.1W
0.001 KW
0.01 KW
0.1 KW
49
Accuracy of Readings
±1% ± 0.8W
±1% ± 8W
±1% ± 80W
±1% ± 0.8KW
±1% ± 8KW
Accuracy of Readings
±1% ± 0.8W
±1% ± 8W
±1% ± 80W
±1% ± 0.8KW
±1% ± 8KW
±1% ± 80KW
Accuracy of Readings
±1% of range
±1% of range
±1% of range
±1% of range
1
2
3
< 20V or < 30A
±2% of range
±2% of range
±2% of range
±2% of range

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pce-6801Pce-6802Pce-3007Pce-830-1

Table of Contents