Figure A-1: Serial Port Pin Layout; Real-Time Printout - Nellcor NPB-190 Service Manual

Pulse oximeter
Table of Contents

Advertisement

Appendix
A4

REAL-TIME PRINTOUT

A-2
The pin layouts are illustrated in Figure A-1. The conductive shell is used as
earth ground. An AMP connector is used to connect to the serial port. Use AMP
connector (AMP p/n 747538-1), ferrule (AMP p/n 1-747579-2) and compatible
pins (AMP p/n 66570-2).

Figure A-1: Serial Port Pin Layout

The serial cable must be shielded. Connectors at both ends of the serial cable
must have the shield terminated to the full 360 degrees of the connector's metal
shell. If rough handling or sharp bends in the cable is anticipated, use a braided
shield.
When a real-time printout is being received, a new line of data is printed every 2
seconds. Every 25th line will be a column heading line. A column heading line
will also be printed any time a value in the column heading line is changed. A
real-time printout is shown below in Figure A-2
NPB-190
Version 1.0.0
Time Tag
%SpO2
123456789
100
123456791
100
123456793
100
123456795
100
123456797
100
123456799
100
123456801
100
123456803
100
123456805
100
123456807
100
123456809
100
123456811
- - -
123456813
- - -
123456815
- - -
123456817
- - -
123456819
- - -
123456821
- - -
123456823
- - -
123456825
- - -
123456827
- - -
123456829
- - -
123456831
- - -
123456833
- - -
NPB-190
Version 1.0.0
Time Tag
%SpO2
123456835
- - -
NPB-190
Version 1.0.0
Time Tag
%SpO2
123456837
79*
123456839
79*
Figure A-2: Real-Time Printout
9
10 11 12 13 14 15
1
2
3
4
5
6
CRC XXXX
SpO2 Limit:
PR (bpm)
PA
Status
120
220
124
220
190
220
190*
220
190*
220
190*
220
190*
220
190*
220
190*
220
190*
220
190*
220
- - -
- - -
SD
- - -
- - -
SD
- - -
- - -
SD
- - -
- - -
SD
- - -
- - -
SD
- - -
- - -
SD
- - -
- - -
PS
- - -
- - -
PS
- - -
- - -
PS
- - -
- - -
PS
- - -
- - -
PS
- - -
- - -
PS
CRC XXXX
SpO2 Limit:
PR (bpm)
PA
Status
- - -
- - -
PS
CRC XXXX
SpO2 Limit:
PR (bpm)
PA
Status
59*
220
59*
- - -
PS SL PL LB
7
8
30-100%
PR Limit: 100-180 bpm
PH
PH
PH
PH
PH
PH LB
PH LB
PH LB
LB
LB
30-100%
PR Limit: 100-180 bpm
80-100%
PR Limit: 100-180 bpm
SL PL LB