Splitting Steps; Matching With A Meter - Accuprint AP30CL - 6000 Service Manual

Closed loop cooling high resolution printer
Table of Contents

Advertisement

3-8

Splitting Steps

A densitometer can be used to assist in determining a fractional difference
in step scales. When measuring density, make several measurements to
avoid being influenced by pinholes.
Example
Pick a step from the reference with a density greater than .15 and less
than 2.0.
Find the steps on the second scale that straddle the first. In the example
the reference is .30, step "A" is .24 and step "B" is .40. (Ref. - A) / (B -
A) = .06 / .16 = .375. The example step is .375 steps under the reference.

Matching with a meter

If an integrating meter is available that has a photocell that matches the
film, it can be used to match the surfaces. We recommend that the results
be verified on film.
Make a reading on the reference surface and then the surface to be
adjusted.
Divide the reading on the surface to be adjusted by the reading from the
reference surface.
Multiply the current calibration number by the result above.
Original
Reference
Second Scale
94MS41 Rev. C1 S/W Ver. 2.06
A
B