Siemens SIWAREX FTC Device Manual

Siemens SIWAREX FTC Device Manual

Module for belt scale
Table of Contents

Advertisement

Quick Links

®
SIWAREX
FTC
Device Manual
Status 10.2010
Module for Belt scale

Advertisement

Table of Contents
loading

Summary of Contents for Siemens SIWAREX FTC

  • Page 1 ® SIWAREX Device Manual Status 10.2010 Module for Belt scale...
  • Page 2: Revision

    Brand names / Trademarks SIWAREX®, SIMATIC®, SIMATIC HMI® and SIMATIC NET® are trade marks of the Siemens AG. The other designations in this info may be trade marks the use of which by third persons for their own purposes might infringe proprietor rights.
  • Page 3 SIWAREX FTC Warning and Safety Terms Contents Preface SIWAREX FTC Scope of Delivery Module for belt scale Product Overview Hardware Configuration and Device Manual Assembly Weighing Functions and Parameters Commands Messages and Diagnosis Programming in SIMATIC STEP 7 Commissioning with a PC –...
  • Page 4: Table Of Contents

    Five Basic Rules for Guaranteeing EMC .............. 4-12 ................4-13 SSEMBLY ON THE ROFILE ..................4-14 ONNECTION AND IRING 4.4.1 Connection Areas for SIWAREX FTC ..............4-14 4.4.2 Shielding Connection .................... 4-15 4.4.3 Connection of 24 V Power Supply ................ 4-16 4.4.4 Connections to Front Terminal ................
  • Page 5 SIWAREX FTC 5.2.7 DR3 - Scale Name ....................5-40 5.2.8 DR3 - Number of Weight Ranges ................. 5-40 5.2.9 DR3 - Scale Type ....................5-40 5.2.10 DR3 - Activate Zero Setting upon Switching On ..........5-40 5.2.11 DR3 - Activated Zero Setting upon Switch On, If Scale Tared ......
  • Page 6 DR 7 - MMC Log Overflow, MMC Trace Overflow, Target Memory for Trace Function ........................ 5-67 5.6.26 DR 7 - Memory Segment for Trace Function ............5-68 5.6.27 DR 7 - Memory Segment for Log ................5-68 viii SIWAREX FTC...
  • Page 7 SIWAREX FTC 5.6.28 DR 7 - Trace Function Recording Cycle ............... 5-68 DR 8 D ....................5-68 DR 9 I ................... 5-69 NFO ON ODULE 5.8.1 DR 9 - Info on Module ................... 5-70 DR 14 T ....................5-70 ILT ANGLE 5.9.1...
  • Page 8 ECHNOLOGY ESSAGES ............7-110 ESSAGE IST OF THE PERATING ESSAGES 8 Programming in SIMATIC STEP 7 ................... 8-112 ......................8-112 ENERAL SIWAREX FTC HW C ................8-112 ONFIG SIWAREX FTC STEP 7 – P ..........8-113 IN THE YCLIC ROGRAM FB SIWA_FTC ..............
  • Page 9 ONDITIONS 14.13 ......................14-133 ICENSES 15 Index ..........................15-134 16 Abbreviations ........................ 16-137 Images FIGURE 3-1 CONFIGURATION OF SIMATIC S7/PCS7 WITH SIWAREX FTC ..... 3-8 FIGURE 3-2 SIWATOOL FTC_B OVERVIEW ..............3-9 FIGURE 4-1 SIWAREX FTC FRONT VIEW ..............4-14 FIGURE 4-2 SHIELD CLAMP ASSEMBLY ..............
  • Page 10 Tables TABLE 1-1 VALIDITY OF THIS MANUAL ..............1-1 TABLE 1-2 OVERVIEW OF CHAPTER ................. 1-2 TABLE 4-1 REQUIREMENTS FOR N SIWAREX FTC MODULES ......4-11 TABLE 4-2 POWER SUPPLY CONNECTION ............. 4-16 TABLE 4-3 LOAD CELL CONNECTIONS ..............4-17...
  • Page 11 SIWAREX FTC TABLE 7-4 LIST OF OPERATING MESSAGES – COMING ........7-111 TABLE 14-1 DATA: POWER SUPPLY 24 V .............. 14-126 TABLE 14-2 DATA: POWER SUPPLY FROM S7 BACKPLANE BUS ...... 14-126 TABLE 14-3 DATA: LOAD CELL CONNECTION ............14-127...
  • Page 13: Preface

    1.2 Basic Knowledge Required To understand this manual, a general knowledge of SIMATIC automation technology is required. Weighing technology is also beneficial. 1.3 Scope of Manual This manual applies to the SIWAREX FTC module: Type Name Order number From product...
  • Page 14: Table 1-2 Overview Of Chapter

    Description of the scope of delivery for Scope of Delivery SIWAREX FTC. Preface Overview of -Structure - Functionality - System integration of SIWAREX FTC. Description Hardware - of individual hardware components Configuration and - of structure and installation Assembly - of connections - of preparations for operation.
  • Page 15: Further Support

    REFACE 1.4 Further Support Do you still have questions concerning the use of the SIWAREX FTC? Then please contact your Siemens representative in the office or business location that is responsible for your area or technical support for SIWAREX Tel.: +49 (0)721 595 2811.
  • Page 16: Scope Of Delivery

    2.1 Scope of Delivery A bus connector for the SIMATIC-bus, the conformity details from the manufacturer and a sheet of additional, current product information belong to the SIWAREX FTC scope of delivery. Configuration packages are required for planning projects with the SIWAREX FTC...
  • Page 17: Product Overview

    SIWAREX FTC (Flexible Technology Continuous Weighing) is a versatile and flexible weighing module. The function module (FM) SIWAREX FTC is integrated in SIMATIC and uses all features of the modern automation system such as integrated communication, diagnostics system and project configuration tools to its advantage.
  • Page 18: Application Range

    3.4 Structure SIWAREX FTC is a function module (FM) of the SIMATIC S7-300 and can be read directly on the SIMATIC S7-300 or ET 200M bus. Installation / cabling efforts for the 80 mm wide module is simplified with the profile rail assembly (snap-in technology).
  • Page 19: System Integration In Simatic

    The SIWAREX FTC has two serial interfaces. The RS 485 interface is used for connecting digital remote displays (for operation as a non-automatic weighing instrument) or force compensation load cells WM, WMH from the Mettler Toledo company.
  • Page 20: Commissioning And Service With Siwatool Ftc_B

    During a service procedures, you can analyze the processes in the scale and test them with the help of a PC. Reading the diagnostics buffer from the SIWAREX FTC is very helpful in analyzing events. Besides complete access to all parameters, memory or print-outs of the weighing file, the program can create weighing curves as well.
  • Page 21: Firmware Download With Siwatool Ftc_B

    The weighing logs are stored on an MMC (Micro Memory Card) that can be inserted into the SIWAREX FTC. If a person disagrees with the results of a certain weighing procedure, the weighing data for that weighing procedure can be reconstructed from the MMC memory.
  • Page 22: Hardware Configuration And Assembly

    Danger Commissioning is not permitted until it is guaranteed that the machine in which these components are to be integrated meets with the guidelines in 89/392/EWG. 4-10 SIWAREX FTC...
  • Page 23: Configuring The Hardware In Simatic

    SSEMBLY 4.1 Configuring the Hardware in SIMATIC SIWAREX FTC is a function module (FM) in the SIMATIC S7 300 automation system. It can be installed in any position designated for function modules. Utilization with the SIMATIC S7 300 is possible in central operation, in an expansion device or decentralized in the ET 200M system.
  • Page 24: Possible Effects Of Interference

    4.2.3 Possible Effects of Interference Electromagnetic interference can influence the automation system and the SIWAREX FTC in various ways: Electromagnetic fields that have a direct influence on the system Interference that infiltrates the environment through bus signals (PROFIBUS DP etc.) Interference through process cabling (e.g.
  • Page 25: Assembly On The Profile Rail

    The shielding of measurement lines must be fastened to ground on both ends. Run cable shielding directly under the SIWAREX FTC on the shielding channeling. The shielding is to be run to the connection terminal.
  • Page 26: Connection And Wiring

    3. Install the shielding strip under SIWAREX. 4. Hang the SIWAREX FTC in its place. 5. Fasten the SIWAREX FTC with 2 screws in the lower area of the module. 6. Label the SIWAREX FTC corresponding with your identification system.
  • Page 27: Shielding Connection

    The following figure shows the assembly of the shield clamps. Figure 4-2 Shield clamp assembly The shielding should be pressed against the shielding rail by the shielding clamp. SIWAREX FTC 4-15...
  • Page 28: Connection Of 24 V Power Supply

    D-sub connector for RS 232. Note In the S7 300 or ET200M system, the SIMATIC CPU or IM 153 and SIWAREX FTC must be connected to the same 24 V supply.
  • Page 29: Table

    Using a junction box is required if more than one LC is connected (the LC must be connected parallel to one-another). If the distance from the LC to the SIWAREX FTC is greater than the existing length of the LC connection cable, then the extension box EB is to be used.
  • Page 30: Digital Inputs

    Load cell connection in 4-wire system Figure 4-4 Load cell connection in 6-wire system 4.4.6 Digital Inputs SIWAREX FTC has 7 digital inputs. These inputs are potentially isolated from the module group. They have a common reference point (M). 4-18 SIWAREX FTC...
  • Page 31: Table 4-4 Digital Input Connections

    SSEMBLY The functionality of each input can be defined with parameters. DI status is indicated with and LED on the front of the SIWAREX FTC. Warning The inputs may only be activated if your functional allocation is known and the activation will not cause any harm.
  • Page 32: Counter Input

    They have a common ground and a secure 24 V supply. They are short-circuit and overload protected. DO status is indicated with and LED on the front of the SIWAREX FTC. When connecting inductive consumers, the digital output that is used must be equipped with a free-wheeling diode.
  • Page 33: Table 4-6 Digital Output Connections

    DO 6 Parameter definable function X1.17 DO 7 Parameter definable function X1.18 DO 8 Parameter definable function X1.19 24 V for digital outputs X1.20 Ground for digital outputs Table 4-6 Digital output connections Figure 4-7 Digital outputs SIWAREX FTC 4-21...
  • Page 34: Analog Output

    Siwarex. The following interface parameters (data record 3 and 7) are to be set: (DR3) Device selection: Digital load cell Mettler Toledo (DR7) Baudrate: 38400 (DR7) Character parity: even (DR7) Data bits: 7 (DR7) Stop bits: 1 4-22 SIWAREX FTC...
  • Page 35: Figure

    Send Stop instruction X1.12 DO 2 (si) Table 4-8 RS 485 Connections Figure 4-9 Connecting Mettler Toledo type WM or WMH to RS 485 Connecting METTLER TOLEDO type WM or WMH some additional parameters must be set. SIWAREX FTC 4-23...
  • Page 36: Rs 485 Interface

    4.4.12 Connecting the Remote Display from Siebert One or more Siebert Type S102 displays can be connected to the RS485 interface. The displays can only be used for displaying weights in "Non automatic weighing instrument" operating mode. 4-24 SIWAREX FTC...
  • Page 37: Connecting The Pc For Siwatool Ftc

    4.4.13 Connecting the PC for SIWATOOL FTC Table RS 232 Interface Termination Clamp Signal Comment 9 pin D-type RS 232 Interface Table 4-10 Connecting the PC Accessories Ready-made cables are available for connecting the PC (see Figure 4-12 Connecting the PC SIWAREX FTC 4-25...
  • Page 38: Led Display Elements

    Approximately 16 MB of data can be Accessories stored with the MMC described in chapter Warning notice After formatting the MMC with SIWAREX FTC, it can no longer be used in the SIMATIC-CPU! Therefore, the MMC should be labeled correspondingly. 4-26 SIWAREX FTC...
  • Page 39: Preparing For Operation

    Is the profile rail connected to the ground conductor? • Have all tools, materials or components that do not belong to the S7 or SIWAREX FTC been removed from the profile rail and the module group? 24 V to Turn on power supply on.
  • Page 40: Application In Ex-Zones

    D A N G E R E X P L O S I O N Exists when the installation and setup regulations are not adhered to. For using the SIWAREX FTC in areas with a risk of explosion, the important information in the “SIMATIC Automation Systems - Basics of Explosion protection (Doc.No.
  • Page 41 Connection from load cells to SIWAREX FTC: The SIWAREX FTC can be connected with load cells in zone 1 or zone 2. For a connection of load cells in zone 2 it is sufficient, that the load cells also have an approval for zone 2 and all associated directives as well as the installation instructions are respected.
  • Page 42: Weighing Functions And Parameters

    5 Weighing Functions and Parameters 5.1 General The SIWAREX FTC can be used as a Non Automatic Weighing Instrument, for force measurement, belt scale, loss-in-weight scale or as bulk flow measurement device. The operating mode is set once when commissioning the scale in the adjustment data area (data record 3).
  • Page 43: Data Record Overview

    The parameters of a data record are shown in a table to start with. This is then followed by the exact parameter description for the parameters of that data record. After the receipt of new parameters, SIWAREX FTC performs a SIWAREX FTC...
  • Page 44: Dr3 Adjustment Parameters

    If there is a parameter error then the data record will not be accepted (stored) by the SIWAREX FTC and a "synchronous" message is generated (see chapter Messages and Diagnosis 5.2 DR3 Adjustment Parameters...
  • Page 45: Table 7-1

    Automatic zero Bit 3: 5.2.11 adjustment 0: Zero tracking device switched off; 1: Zero tracking device switched on Bit 4 - 7 not set 5.2.12 (2) Calibration parameter Minimum weight for REAL DBD058 5.2.13 Weighing range 1 SIWAREX FTC 5-33...
  • Page 46 4% for country code „OIML“) Tare max. weight T- BYTE DBB110 Subtractive tare device range 5.2.29 [in % of WB with multi-range scale] [in % of WB with multi-resolution scale] (Entry may not exceed 100% with country code "OIML") 5-34 SIWAREX FTC...
  • Page 47: Dr3 - Adjustment Digits 0, 1, 2, 3, 4, For The Zero Point And Adjustment Weights 1, 2, 3, 4

    The analog measurement value coming from the load cells is converted into a digital value with an analog-digital converter. A weight value is then calculated from this digital value. All functions of the SIWAREX FTC then uses this weight value for executing their task.
  • Page 48 (without the weight of the construction itself) is 1 398 101 the plausibility of the characteristic curve can be roughly estimated to determine load deviations for example. The illustration shows the relationship between the adjustment digits and the adjustment weight of the scale. 5-36 SIWAREX FTC...
  • Page 49: Technical Data

    FIGURE 5-1 Adjustment Digits and Weight Value If the adjustment weights and adjustment digits of a SIWAREX FTC are known then the adjustment procedure does not need to be performed. They are simply sent to the SIWAREX FTC in the DR3 data record and the scale is immediately operational (after official calibration of the scale, it is no longer possible to send DR3).
  • Page 50 Linearization of the scale characteristic curve The characteristic curve is converted for bi-directional operation in "Force measurement" operating mode. The following image shows the relationship between effective force and the adjustment digits. FIGURE 5-3 Adjustment Digits and Force 5-38 SIWAREX FTC...
  • Page 51: Dr3 - Characteristic Value Range

    Depending on the characteristic value of the connected load cells, the value 1 mV/V, 2 mV/V or 4 mV/V must be selected. Since the SIWAREX FTC supplies the load cells with approximately 10 V, the measuring input is redefined according to the expected measurement voltage (max.
  • Page 52: Dr3 - Limit Frequency

    The weight value is based on the average of the n weight values that the SIWAREX FTC calculates every 10 msec, e.g. if n=10, 10 values are used for calculating the average value. Every 10 ms, the oldest value drops out of the calculation and the newest is considered in the calculation.
  • Page 53: Dr3 - Activated Zero Setting Upon Switch On, If Scale Tared

    The current weight can be used for calibrating purposes (with the numeral step for measurement range 1) only under the maximum weight. The maximum weight is defined during adjustment. The maximum weight depends on the number and type of load cells used and the required weight range. SIWAREX FTC 5-41...
  • Page 54: Dr3 - Minimum Weight For Weighing Range 1

    3) only under the maximum weight. The maximum weight is defined during adjustment. The maximum weight depends on the number and type of load cells used. Note If only one or two weighing ranges are defined then this entry is meaningless. 5-42 SIWAREX FTC...
  • Page 55: Dr 3 - Numeral Step For Weighing Range 3

    To prevent having to cancel a weighing instruction (only taring, printing and zero setting in static operation) immediately when the stable weight of the scale is not defined, the waiting time is given as standstill. If the scale instruction tare, output SIWAREX FTC 5-43...
  • Page 56: Dr 3 - Maximum Negative Weight For Zero Setting Upon Switch On

    4% of the maximum weighing range. 5.2.29 DR 3 - Tare max. Weight T- SIWAREX FTC accepts every weight value which is less than %-record of the maximum weighing range with multi-range scales or weighing range 1 with a multi- resolution scale.
  • Page 57: Dr 3 - Weighing Operating Mode: Nawi Filling Procedure

    Data records or individual parameters, which have no effect in a certain operating mode, are also taken and stored in the SIWAREX FTC module. The following operating modes can be defined: 0 - non-automatic weighing instrument (OIML R 76) for fill scales...
  • Page 58: Dr 3 - Load Cell Type

    WMH is monitored. If no telegram is received for the defined time period (in ms), SIWAREX FTC indicates an operational fault No. 24 – Digital load cell not ready. A status bit also provides (Status NAWI, Bit 29) the information on whether the digital load cell is ready.
  • Page 59: Dr 3 - Regulations

    The determination time applies for the following instructions: zero point valid, adjustment weight 1...4 valid, zero setting, taring. Example: SIWAREX FTC 5-47...
  • Page 60: Theoretical Adjustment - Adjustment Without Adjustment Weights

    DBD004 2000 Time within which, a recording task must be 5.4.1 recording ended (msec) Device for log output BYTE DBB008 Bit 0 : 5.4.2 0 = Log output to RS232 interface 1 = Log output to MMC 5-48 SIWAREX FTC...
  • Page 61: Dr 4 - Monitoring Time For Logging

    DR 4 - Monitoring Time for Logging If logging of the weighing results is generated by an instruction or automatically, SIWAREX FTC monitors whether the procedure has been completed within the predefined time. If not, an error is generated. 5.4.2 DR 4 - Device for the Log Output Logs can be output to a connected printer or to the Micro Memory Card (MMC).
  • Page 62: Dr 4 - Basis Weight For Monitoring The Empty Range

    The value for the empty range is a limit value, that when exceeded indicates the "empty" status to the SIWAREX FTC which gives this out as status information. In operating mode Loss-in-weight scale, emptying in automatic mode is ended if the value drops below the limit value.
  • Page 63: Dr 4 - Cut-Off Weight Limit Value 2

    Correction factor for the flow rate 5.5.11 flow-rate Minimum flow-rate DBD044 Limit value for minimum flow-rate in % of the 5.5.12 nominal flow-rate Maximum flow-rate DBD046 Limit value for maximum flow-rate in % of the 5.5.13 nominal flow-rate SIWAREX FTC 5-51...
  • Page 64: Dr 5 - Nominal Speed

    DR 5 – Time basis for the belt speed The belt speed is output in LU (length units) per second. The measuring time of the pulse input can be set to another time e.g. 2 seconds. That 5-52 SIWAREX FTC...
  • Page 65: Dr 5 - Pulse Constant

    Minimum belt speed is a limit value for minimum belt speed in % of the nominal speed. The entry of the number e.g. 800 corresponds with 80.0 %. After this value has been undercut, SIWAREX FTC sets a status bit (see scale status...).
  • Page 66: Dr 5 - Minimum Flow-Rate

    Minimum belt loading for totalizing is a limit value for minimum belt loading in % of the nominal belt loading. Entering the number e.g. 20 corresponds with 2.0 %. After the actual value is less than the limit, SIWAREX FTC will not execute the totalizing function.
  • Page 67: Dr 5 - Numeral Step For Totals 1

    For the external quantity counter, the pulse duration can be defined so that the device can recognize the pulse for sure. The pulse duration is not permitted to be selected so long that stopping the minimum pause duration will make the next pulse late. SIWAREX FTC 5-55...
  • Page 68: Dr 5 - Over And Under-Load Inhibition Time

    If the status remains for a longer period of time, the fault message is output. 5.6 DR 7 Interfaces The parameters that define the behavior of the SIWAREX FTC on the interfaces are defined in DR 7. If an interface is not used the default values can be left alone.
  • Page 69 Value range 100h – 13Fh 100h+ 0...1Fh Bit-No. NAWI-status bit - coming 100h+ 20h...3Fh Bit-No. Conti status bit-coming Value range 200h – 23Fh 200h+ 0...1Fh Bit-No. NAWI status bit - going 200h+ 20h...3Fh Bit-Nr. Conti status bit-going SIWAREX FTC 5-57...
  • Page 70 0: 0 .. 20mA 1: 4 .. 20mA Bit 1 – 7 unused RS232 RS232-printer baud rate BYTE DBB042 0 = 1200 Bit/sec. 5.6.13 1 = 2400 Bit/sec. 2 = 4800 Bit/s 3 = 9600 Bit/s Other definitions not permitted. 5-58 SIWAREX FTC...
  • Page 71 20h...3Fh Bit-No. Conti status flag FFh output always inactive Definition of the digital BYTE DBB051 0FFh Value range 00h – 3Fh 5.6.19 output 4 0...1Fh Bit-No. NAWI status flag 20h...3Fh Bit-No. Conti status flag FFh output always inactive SIWAREX FTC 5-59...
  • Page 72 Bit 0: Digital input 1 -> 0 = high active: 1 = low 5.6.24 inputs 1...7 active ..Bit 6: Digital input 7 -> 0 = high active: 1 = low active Bit 7: Not used Reserved TIME DBD068 Reserved 5-60 SIWAREX FTC...
  • Page 73: Dr 7 - Source For The Weight Simulation

    5.6.3 DR 7 - Force in Service Operation After activating this function, the digital outputs can be forced independent of their parameter definitions. Force control is only possible in service operation and only with the SIMATIC interface. SIWAREX FTC 5-61...
  • Page 74: Dr 7 - Process Value 1 For Fast Output To The Simatic Cpu

    SIWAREX FTC, the peripheral interface of the SIWAREX FTC exists. The data is read by the FB43 cyclically and provided for the user as an output variable. In the belt scale operating mode, normally the current belt load (selection code 45) is of some importance.
  • Page 75: Dr 7 - S7-Fb-Life Bit Monitoring Time

    5.6.7 DR 7 - S7-FB-Life Bit Monitoring Time The SIWAREX FTC can determine whether the FB is maintaining the communication with the SIMATIC CPU by monitoring the life bit. If the FB does not respond, an operating error is output after the defined time has elapsed. Monitoring is active in RUN status of the SIMATIC_CPU.
  • Page 76: Dr 7 - Rs232 Printer Baudrate

    5.6.17 DR 7 – RS 485 Baudrate The following protocols can be set for the RS485 interface: 0 = 1200 Bit/sec. 1 = 2400 Bit/sec. 2 = 4800 Bit/s 3 = 9600 Bit/s 4 = 19200 Bit/s 5-64 SIWAREX FTC...
  • Page 77: Dr 7 - Rs485 Character Frame

    If bit 0 is 0 then DO1 is high active and if bit 0 is 1 then DO1 is low active, If bit 1 is 0 then DO2 is high active and if bit 1 is 1 then DO2 is low active, etc. SIWAREX FTC 5-65...
  • Page 78: Dr 7 - Replacement Values For Do 1 To 8 With Fault Or Output Disable

    If bit 1 is 1 then DO2 is 1 with fault. etc. Warning If an output is to be set with the fault (operating error) then you must ensure that it does not lead to a hazardous situation. 5-66 SIWAREX FTC...
  • Page 79: Dr 7 - Definition Of Digital Inputs 0, 1, 2, 3, 4, 5, 6, 7

    The trace function is described in Bit 2 defines which memory is used for recording the trace data. 0: Trace data is stored in RAM 1: Trace data is stored in the MMC 5.6.28 The trace function is described in SIWAREX FTC 5-67...
  • Page 80: Dr 7 - Memory Segment For Trace Function

    5.7 DR 8 Date / Time SIWAREX FTC has its own hardware clock. The actual date and time can be read through DR 8. The date and time can be set on the module group using this data record if required.
  • Page 81: Dr 9 Info On Module

    Reference Info on module CRC Checksum DWORD DBW000 Length of firmware in DWORD DBW004 bytes Product Info STRING DBB008 Siemens AG Serial number of the module [26] Module name STRING DBB036 SIWAREX [10] Application STRING DBB048 „CONT“ Application identification [4][8] „NAWI“...
  • Page 82: Dr 9 - Info On Module

    Allocation of DR 9 5.8.1 DR 9 - Info on Module The information on the SIWAREX FTC module is for the purpose of identifying the module in the manufacturer's plant (e.g. for repairs). This information has no influence on user operation.
  • Page 83: Dr 15 - Tare Entry

    If the DR 16 was defined as the source for the weight simulation (see DR 7 for the weight simulation), defining a weight value through DR 16 will deactivate the measurement input of the SIWAREX FTC and the defined value is "simulated" as the weight value. Procedures:...
  • Page 84: Dr 17 - Ext. Definition For Analog Output

    Send DR 21 to the scale Name Type Address Default Range of Values / Meaning Reference Load set value REAL DBD000 Quantity to be dosed in batch operation 5.14.1 in reference to WU, Switch off in reference with totalizing memory 1 5-72 SIWAREX FTC...
  • Page 85: Dr 21 - Loading Quantity

    The status can also be monitored if the SIWAREX FTC is controlled by the SIMATIC CPU. The NAWI- and Conti-status flags are simultaneously updated in DR 32 as well - process status loss-in-weight scale.
  • Page 86: Dr 30 - Nawi Status Bits

    The MMC is in position MMC ready The MMC is formatted and ready for recording MMC-trace ready The MMC is ready for the trace function MMC-log ready The MMC is ready for logging Trace active The trace function is activated 5-74 SIWAREX FTC...
  • Page 87: Dr 30 - Conti-Status Flags

    The maximum filling time has been achieved. Max. time max. control The maximum control difference is at the max. defined difference monitoring time. Max. vol. time The volumetric mode is active longer than the defined max. monitoring time. Reserved Reserved SIWAREX FTC 5-75...
  • Page 88: Dr 30 - Gross Process Value

    The current belt speed in % of the predefined nominal speed. 5.15.11 DR 30 – Belt Loading in % of the Nominal Load The current belt load is output as the relationship of the current belt load to the nominal belt load. 5-76 SIWAREX FTC...
  • Page 89: Dr 30 - Nominal Load

    The current flowrate is output in % of the nominal flowrate. 5.15.18 DR 30 – Operating Hours Counter The operating hours counter records the number of hours that the SIWAREX FTC is supplied with 24V voltage. 5.15.19 DR 30 – Pulse Counter Value Current counter value on the pulse input in relation to the defined measurement time.
  • Page 90: Dr 31 - Unfiltered Digital Value

    The current existing operating errors (fault) are output as double word. 5.16.4 DR 31 – Date Time The current date and time from the internal clock of the SIWAREX FTC. 5.16.5 DR 31 – Max. Temperature The maximum occurring module temperature (slave pointer).
  • Page 91: Dr 33 - Distribution Memory

    „Delete distribution memory 5“ or „Log and delete distribution memory 5“. 5.17.6 DR 33 – Distribution Memory 6 Content of distribution memory. Dimension in Large weight unit [LWU]. Can be reset with instruction „Delete distribution memory 6“ or „Log and delete distribution memory 6“. SIWAREX FTC 5-79...
  • Page 92: Dr 34 Ascii Display Value

    Table 5-18 Allocation of DR 34 SIWAREX FTC controls the content of DR 34 via instructions 25, 26, 128...142, 181… 192. 5.19 DR 35 Encoded Information for Verifiable Display The content of the DR 35 corresponds with the DR 34, but is encoded. The verifiable display is generated on the SIMATIC OP/TP from the contents of DR 35.
  • Page 93: Table 5-12 Allocation Of Dr

    Log text entries 1 to 4 are composed of fixed segments such as title and label and of variable fields. The variable fields can contain various values from the SIWAREX FTC. In addition, 4 designation variables can be defined in addition to the process variables.
  • Page 94 BB 1234567.8kg/m Flow in % FS•123•%•••••••• Belt load in % BB•123•%•••••••• Belt speed in % BS•123•%•••••••• Current setpoint SP•1234567.8kg/s Effective setpoint SW•1234567.8kg/s Current control difference DI•1234567.8kg/s Act. Control difference in % D•123,00•% Act. Difference vol/grav in % K•123,00•% 5-82 SIWAREX FTC...
  • Page 95: Dr 44 Last Log

    The last output log data can be read up until the next logging procedure. 5.22 DR 45 String The string is test that can be inserted as variables in the log. The content of the string can be defined through SIWATOOL or from control program of the SIMATIC CPU. SIWAREX FTC 5-83...
  • Page 96: Dr 46 Parameter For Reading Mmc Logs In Simatic

    The user enters the desired log ID in DR 46 and then sends DR 46 to the SIWAREX FTC. A short time later, the log is prepared for reading (Status bit „MMC data ready S7). If the user now requests the DR 47 with the log, the log with the desired ID is read in DR 47.
  • Page 97: Dr 47 Requested Log

    The trace function is used for logging measurement values and current states of the scale in RAM memory of the SIWAREX FTC or in the MMC card. When logging to RAM, SIWATOOL FTC reads the trace elements with data record 120 and when logging to the MMC it is done through data record 121.
  • Page 98: Dr 123 Data Content Mmc

    An element requires 46 bytes. If the recording is run e.g. once per 50 msec, it means memory requirements of 920 bytes per second. In the simplest case, the RAM memory (approx. 2 MB) of the SIWAREX FTC can be used for the recording.
  • Page 99: Dr 122 Log Data Mmc

    0, the requested log number does not exist Log ID DINT ID for stored log text Log text BYTE[n] Log text 1|2|3|4 Block check character CRC16 (MMC-ID + Log ID + Log text[n]) Table 5-28 MMC log SIWAREX FTC 5-87...
  • Page 100: Commands

    Every command has a unique number. A command can be sent via various interfaces (SIMATIC, SIWATOOL FTC, digital input). Every time that a command is sent to SIWAREX FTC, a check must be made for whether the command has been executed correctly. The data and operating errors (synchronous errors) that are generated provide information on why the command could not be executed.
  • Page 101: Command List

    Adjustment instruction Adjustment weight 4 Service operation valid The fourth adjustment weight is assigned with the momentary weight. Predefine all data records with default values Service operation All parameters are set to the status that was assigned originally by the manufacturer. SIWAREX FTC 6-89...
  • Page 102 The defined tare is accepted as tare and at the same time, "Preset tare" is designated together with the weight display. Switch on increased resolution Activate output / display of the verifiable weight value with increased resolution for 5 seconds. 6-90 SIWAREX FTC...
  • Page 103 Output the log with text layout 4 Repeat last logging procedure The last log output is repeated. 40-41 Reserved Reserved Reserved Micro Memory Commands Start the Recording The recording (trace function) should start End running recording The activated recording (trace function) is ended. SIWAREX FTC 6-91...
  • Page 104 This is used for stopping the totalizing, the flowrate continues to be measured and there is no integration into the distribution memory. Loading operation is ended. Switch on dosing (belt scale) Activate discharge organ (is only used if the belt control is switched on) 6-92 SIWAREX FTC...
  • Page 105 Log and delete distribution memory 3 Deletes distribution memory 3. This automatically includes a logging procedure. Text 1 (DR 40) is used here. The deletion is only executed if total 3 is contained in the log text. SIWAREX FTC 6-93...
  • Page 106 Deletes verifiable distribution memory 4 This automatically includes a logging procedure. Text 1 (DR 40) is used here. The deletion is only executed if total 1 is contained in the log text. Delete distribution memory 4 Deletes distribution memory 4. 6-94 SIWAREX FTC...
  • Page 107 Display total weight value in DR 34 (total 1 from DR 33). Display distribution memory 2 Display total weight value in DR 34 (total 2 from DR 33). Display distribution memory 3 Display total weight value in DR 34 (total 3 from DR 33). SIWAREX FTC 6-95...
  • Page 108: Table 6-1 Siwarex Ftc Command List

    Display of the belt speed (% of the nominal speed of the belt) in DR 34 (from DR 31). Table 6-1 SIWAREX FTC command list The commands in the table above can be activated over all interfaces. 6-96 SIWAREX FTC...
  • Page 109 Range for combined commands. The function block SIWA_FTC (FB44) can transfer several data records in sequence. Read all data records of belt scale in the SIWAREX FTC (DR3, DR4, DR7, DR5, DR8, DR9, DR14, DR15, DR16, DR17, DR18, DR20, DR21, DR30, DR31, DR32, DR34, DR35, DR44, DR45, DR47).
  • Page 110: Messages And Diagnosis

    7.2 Message Paths The SIWAREX FTC messages are passed on to the user by various paths. During configuration, it is important to choose the right path for routing and processing the messages.
  • Page 111: Detection Of Messages Using Siwatool Ftc_B

    For more information see chapter STEP 7 Recognizing messages with process alarms. Process alarms enable the user to respond flexibly to technology messages or to 8 Programming the selected status information. For more information see chapter in SIMATIC STEP 7 SIWAREX FTC 7-99...
  • Page 112: Message List Data And Operating Errors

    Attention! A pre-calibrated scale will lose its calibration. Transmission of the The data record or the command is not permitted to be data record or activated during the totalizing. command is not permitted in continuous operation 7-100 SIWAREX FTC...
  • Page 113 Trace task cannot be executed, since MMC is defective or non- existent Trace recording cycle The recording cycle is too small in DR 7. Define a longer too small recording cycle (at least 5 x 10 msec. for MMC). SIWAREX FTC 7-101...
  • Page 114 4% of the maximum weighing range, or > 20% of the zero setting range in DR 3 in verifiable operation (technical regulation code OIML in DR 3). The sum of the negative and positive values apply in this instance. Reduce the zero definition range or zero setting range. 7-102 SIWAREX FTC...
  • Page 115 Adjustment weight too The adjustment weight is too low in DR 3. Increase the adjustment weight interval. The measurement values between two sequential adjustment weights must be separated by a space of at least 2% of the measurement range. SIWAREX FTC 7-103...
  • Page 116 (must be larger than 0). Allocation of log text The log text number in DR 21 is incorrect. Correct the number No. for autom. logging (0..4). not permitted Belt length Effective belt length implausible implausible (must be larger than 0). 7-104 SIWAREX FTC...
  • Page 117 The numeral range for a definition is invalid. Permitted values weighing values are -100 000 000 and lower than 100 000 000. exceeded Checksum error Checksum error in data record 26. The data was falsified or the DR versions to not match. SIWAREX FTC 7-105...
  • Page 118 10% to the reference value or if an error has occurred during the measurement. DR-writing not The selected data record cannot be defined if the write possible protection is set. Table 7-1 List of Data and Operating Errors 7-106 SIWAREX FTC...
  • Page 119: Message List Technology Messages

    Check the log texts in DR 40, 41, 42 or Timeout print-out A print-job was not completed within the monitoring time defined for logging in DR 4. Check the output device or increase the monitoring time. SIWAREX FTC 7-107...
  • Page 120 Belt load too small, check the material feed. Belt load too large Belt load too large, check the material feed. Flowrate too small Flowrate too small, check the material feed. Flowrate too large Flowrate too large, check the material feed. 7-108 SIWAREX FTC...
  • Page 121: Table 7-2 List Of Technology Messages

    Batch time exceeded The time for one batch (DR21) is exceeded. Impulse output error Overrun of the impulse output buffer. Weighing stopped by SIMATIC-CPU- Weighing stopped by SIMATIC-CPU-stop stop Table 7-2 List of Technology Messages SIWAREX FTC 7-109...
  • Page 122: Message List Of The Operating Messages

    FB SIWA is damaged. Belt speed too low The pulse input no longer delivers pulses (less than 2 pulse/s) Fault on RS485 Communication fault on the RS485 interface. Table 7-3 List of Operating Messages –going 7-110 SIWAREX FTC...
  • Page 123: Table 7-4 List Of Operating Messages - Coming

    FB SIWA is damaged. 151 Belt speed too low The pulse input no longer delivers pulses (less than 2 pulse/s) 152 Fault on RS485 Communication fault on the RS485 interface. Table 7-4 List of Operating Messages – coming SIWAREX FTC 7-111...
  • Page 124: Programming In Simatic Step 7

    Hardware Configuration. SIWAREX FTC is projected as a function module in the SIMATIC manager. In order for the SIWAREX FTC to be accepted into the module assembly catalogue of the SIMATIC manager, the SETUP program (SETUP_FOR_SIMATIC on the project planning package CD) or hardware update by HSP (hardware support package) must be executed.
  • Page 125: Siwarex Ftc In The Cyclic Step 7 - Program

    SIWA_ FTC. While programming the call, an instance data block is created for the FB SIWA_FTC. Besides the instance data block, a scale DB is required for every SIWAREX FTC scale, in which the scale parameters are stored. The UDT that comes with it can also be used for creating the scale DB.
  • Page 126: Db_Scale:= 18, Input, Int

    A scale DB must be defined for every scale, in which the parameter of the SIWAREX FTC and the current actual value are found. The number of the DB can be freely selected. DB15 has been defined as the scale DB in the project planning package.
  • Page 127: Cmd_Err:= "Db_Scale".Bo_Cmd_Err, Output, Bool

    FB SIWA_FTC are read cyclically by the FB through the peripheral range. SIWAREX FTC updates the values internally in a 10 msec rhythm. Every update is assigned a number which can be used in the SIMATIC CPU like a time stamp.
  • Page 128: Err_Msg_C:= "Db_Scale".B_Err_Msg_Code, Output, Byte

    Bit 2 - Error interpreting a data record / command, specified data record or command number is incorrect. Bit 3 - Life bit error, SIWAREX FTC not responding Bit 4 - Unable to read peripheral data in this cycle Bit 5 - Activated command aborted at restart...
  • Page 129 8.4.21 START_UP:= "DB_SCALE".bo_START_UP_IN_PROGRESS Communication between the SIWAREX FTC and the FB SIWA_FTC is synchronized when the SIWAREX FTC module is restarted (normally when the SIMATIC CPU is started). The bit can be output for more than one cycle. 8.4.22 CMD_EN:= "DB_SCALE".bo_CMD_ENABLE After entering the command number in the CMD_IN:= "DB_SCALE".i_CMD_INPUT...
  • Page 130: Commissioning With A Pc - Siwatool Ftc_B

    9.2 Windows and Functions in SIWATOOL FTC_B The program window is constructed so that navigating through the parameter definition areas of the SIWAREX FTC is made easier. In the left-hand portion, the overview of the parameters is shown as a tree-structure. The parameter grouping corresponds with the various activities which can happen during the project planning, commissioning, tests and for service cases.
  • Page 131: Help

    Tips can also be helpful in remembering the meanings of certain parameters (text is displayed if the mouse pointer is positioned over a field). After clicking on menu point "Help", the "SIWAREX FTC" manual is called. Installation of Acrobat Reader is necessary to read the manual.
  • Page 132: Firmware-Update With Siwatool Ftc_B

    The new firmware can slightly vary from earlier versions with small changes - this is the case if the data structure of SIWAREX FTC parameters are not changed. In this case, putting the new firmware on will not change the current data.
  • Page 133 Downloading the Firmware with SIWATOOL FTC_B Warning notice SIWAREX FTC will not respond to the communication requests from the SIMATIC CPU while transferring the firmware. The CPU indicates a peripheral access error and unless a corresponding OB (Organization Block) is programmed, the CPU can go to STOP status.
  • Page 134: Empty

    11 Empty 11-122 SIWAREX FTC...
  • Page 135: Project Planning In Simatic Pcs 7

    SIMATIC PCS 7 ROJECT PLANNING IN 12 Project planning in SIMATIC PCS 7 The integration of SIWAREX FTA is possible as of PCS 7 version 6.1. For description please refer to the separate manual under www.siemens.com/weighing-technology. SIWAREX FTC 12-123...
  • Page 136: Accessories

    SIWAREX FTC Device Manual (The device manual is only available electronically and is available through Internet page www.siemens.de/siwarex) SIWAREX FTC configuration package for SIMATIC S7 for belt scale on CD- 7MH4900-3AK01  SETUP for S7 integration as of Step7 V5.2 ...
  • Page 137: C Onnection And

    Cable Li2Y 1 x 2 x 0.75 ST + 2 x (2 x 0.34 ST) – CY 7MH4702-8AG for connecting a SIWAREX FTC with connection and distribution box (JB), expansion box (EB) or Ex-Interface (Ex-I) as well as between two JBs, for stationary cabling, occasional bending is possible, 10.8 mm outer diameter for...
  • Page 138: Technical Data

    Max. distance from load cell to Ex-i-Interface in 150/500 m for gas group IIC Ex range 1000 m for gas group IIB (see SIWAREX IS device manual) Lowest permitted input signal for a calibration value ≥ 0.5 µV/e in calibrating operation 14-126 SIWAREX FTC...
  • Page 139: Analog Output

    Voltage range for H signal DC 15 V to 30 V Voltage range for L signal DC -3 V to Input current (15 to 30 V) 2 to 15 mA Switching frequency max. 50 Hz max. 50 Hz SIWAREX FTC 14-127...
  • Page 140: Counter Input Ci (For Belt Scale)

    7 or 8 Parity even / odd Stop bits 1 or 2 Max. distance 1000m at 1200 Baud Signal level Conforms with EIA-RS485 Termination resistance 390Ω / 220Ω / 390Ω Isolation 500 V Table 14-8 Data: RS 485 interface 14-128 SIWAREX FTC...
  • Page 141: Dimension And Weight

    , Half sinus EC 60721-3-2, class 2M2 Duration: 6 msec IEC 61131-2 Number: 1000 per axis each DIN IEC 60000-2-32 Test Ed Falling High: 1 m IEC 61131-2 Number of trails: 5 Shipping package Table 14-10 Data: Mechanical Requirements SIWAREX FTC 14-129...
  • Page 142: Electrical-, Emc- And Climatic Requirements

    Protection class I, with IEC 536 protective conductor IP protection DIN 60529 in S7 frame: IP20 IEC 529 only SIWAREX FTC: IP10 Air and creepage distances IEC 1131 Surge category II UL508 Pollution degree 2 CSA C22.2 No.142 Circuit board material IIIa Circuit track spacing 0.5 mm...
  • Page 143 An external protection element should be fitted to meet the requirements (e. g.: Blitzductor VT AD24V, from Dehn&Söhne) Notice This is a class A device. In residential areas, this device can cause radio interference. In such areas, users must take suitable measures (e.g.: use in 8MC cabinets). SIWAREX FTC 14-131...
  • Page 144: Nvironmental Onditions

    14.12 Environmental Conditions The SIWAREX FTC is suitable for weather protected, stationary application in SIMATIC S7-300 systems. Conditions for application conforming with IEC 1131-2. For use in extreme operating conditions (e.g. heavy dust, acidic moisture or gasses etc.), additional measures are to be taken such as B. Encapsulation.
  • Page 145 ECHNICAL 14.13 Licenses Note: The licenses that are currently valid for SIWAREX FTC are indicated on the SIWAREX FTC rating plate. Directives: 2004/108/EC “Electromagnetic Compatibility” 94/9/EG “Explosion protection” (ATEX 100a) Note : Further information on the EC directives can be found in the product documentation accompanying every SIWAREX FTC.
  • Page 146: Analog Output

    Force measurement 5-45 Commissioning, Preparation 4-27 Function Configuration package Further Support Configuring 4-11 Gross process value 5-76 Connection 4-14 Hardware Configuration v, 1-2, 4-10 Connection Areas 4-14 HW Config 8-112 Constant belt speed 5-53 Impedance actual value 5-78 15-134 SIWAREX FTC...
  • Page 147 5-54 Scale type , 5-40 5-33 Minimum belt loading 5-54 Scope of Delivery Minimum belt speed 5-53 Service Minimum flow-rate 5-54 Service and adjustment commands6-88, 6-89 Minimum Weight for Weighing Setpoint 5-72 Range 5-41 Shield connection 4-15 SIWAREX FTC 15-135...
  • Page 148 Tare max. weight T- , 5-44, 5-46 Weight simulation 5-61, 5-71 5-34 Tare process value 5-76 Wiring 4-14 Technical Data 14-156 Zero adjustment , 5-41 5-33 Technology messages 7-107 Zero setting , 5-40, 5-44 5-33 Temperature 5-78 Zero Setting 5-44 15-136 SIWAREX FTC...
  • Page 149 Physikalisch Technische Bundesanstalt (certification authority for calibration of scales) Random Access Memory S7-300 Siemens Automation System for the mid-performance-range S7-400 Siemens Automation System for the high-level performance range System Function Call (S7) STEP 7 Programming device software for SIMATIC S7 Tare weight Totally Integrated Automation...