NOTE:
After recalibration the display will show the
operating mode symbol followed with letters CAL to
indicate that the mode and the volume in question is
recalibrated to give out the new volume.
NOTE:
Changing the mode will reset the recalibration.
However, it is possible to store the recalibrated mode
to the storage location in PROG mode as described in
paragraph 8.

13. Troubleshooting

The eLINE® pipettes have a onboard monitoring
program to control the performance of each pipetting
action. If any error message appears to the display it
means that the pipette has been unable to perform
the attempted action properly. In case of an error the
display will show Error and PRESS Ent. message. To
clear the message please follow these steps:
1. As this procedure will reset the pipette and the tip
ejector please first remove the tip by hand.
2. Recharge the pipette for 15 minutes.
3. Clear the error message by pressing
display will show RESET and PRESS TIP.
3. Reset the pipette by pressing either of the tip
ejector buttons
NOTE:
Repeated occurrence of error messages
indicates an internal error causing failure of the
eLINE® to complete the execution of the pipetting. The
eLINE® therefore needs to be returned to the nearest
Sartorius Service Centre or your local Sartorius service
representative for repair.
.
. The
37