Store Operating Mode To Storage Locations; Recall Stored Mode From The Storage Location; Pipetting Recommendations - Sartorius eLINE User Manual

Electronic pipette
Table of Contents

Advertisement

NOTE:
This procedure overwrites the possible
previously stored mode from the selected storage
location.
9.2. Recall stored mode from the
storage location
To recall a stored mode from Prog 1–6:
1. Press
to display PROG
2. Scroll with
is displayed (1–6)
NOTE:
The display shows simultaneously the mode
symbol of the operating mode stored in to the storage
location in question.
3. Confirm by pressing
4. The pipette is ready to perform according to the
stored mode. Position the tip to aspirate and press
START button
10. Pipetting
Recommendations
The wide selection of different operating modes
makes it possible to perform several different liquid
handling tasks with eLINE® pipettes. The operating
modes P, P+MIXING, P COUNT, dd, dd+MIXING and
SA feature an automatic blow-out and others leave
an excess liquid in the tip. The user should follow
the recommendations below to ensure optimal
performance.
10.1. Dispensing With Blow-out
The automatic blow-out function, in different
pipetting and diluting modes, is followed by an
immediate return of the piston to the home position.
To avoid accidental aspiration of the liquid back into
the tip, it is recommended that the dispensing is
always performed above the liquid surface.
until the desired storage location
.
27