Reverse Pipetting (Rp) - Sartorius eLINE User Manual

Electronic pipette
Table of Contents

Advertisement

To mix:
1. Position the tip in to the solution, then press and
hold the START button
Mixing is performed automatically as long as the
START button
2. Position the tip to dispense. Press START button
twice
NOTE:
The mixing is performed with about 70% of the
total volume.

7.2. Reverse Pipetting (rP)

A selected volume plus and excess is aspirated into the
tip. After dispensing the selected volume the excess
volume remains in the tip and is discarded.
1. Press
to display PIPET
2. Scroll with
3. Confirm by pressing
4. Select the desired pipetting volume by using
NOTE: When holding down
starts to scroll faster.
5. Press
to confirm selection and to display the
speed in
or press
and continue from Step 10.
6. Select the speed in with
7. Press
or
display the speed out
8. Select the speed out with
Max)
9. Press
or
10. Position the tip to aspirate and press the START
button
11. Position the tip to dispense. Press the START
button
16
.
is held down.
until the symbol rP is displayed.
the volume display
(1= Min and 9 = Max)
to confirm selection and to
(1= Min and 9 =
to confirm selections