Trouble; Troubleshooting - Sartorius Picus 2 Operating Instructions Manual

Electronic pipette
Table of Contents

Advertisement

Trouble

10 Trouble
10.1

Troubleshooting

Fault
Droplets remain in the
tip.
Inaccurate function
The device is switched
off.
The piston jams
The aspirated volume is
too small
The device does not
respond.
48
Picus
2 Operating Instructions
®
Cause
The tip is not compatible.
The device is
contaminated.
The device is defective.
The device is
contaminated.
The device is defective.
Pipette in Off mode
The battery is empty.
Internal components are
loose
Pipette contaminated
Pipette defective
The Safe-Cone Filter is
contaminated
The device is
contaminated.
The device is defective
Remedy
Use original Sartorius tips.
Check that the tip is securely seated and,
if necessary, tighten.
Clean the device.
Replace the defective parts.
Send the device in for servicing.
Clean the device.
Replace the defective parts.
Send the device in for servicing.
Switch on the device.
Connect the device to the charger.
Open the lower assembly and ensure the
parts are securely fitted.
Clean the pipette.
Replace defective parts.
Send the device in for servicing.
Replace the Safe-Cone Filter.
Clean the device.
Replace the defective parts.
Send the device in for servicing.
Reset the device to the defaults.
Send the device in for servicing.
Chapter,
page
9, 41
9, 41
6.3.1, 30
6.2, 29
9.2.2, 42
9, 41
9.2.1, 42
5.4, 28
9, 41
10.2, 49