11.3. Battery Replacement

The design of the eLINE® pipettes enables fast and easy
battery replacement. The battery replacement pack
includes 4 NiMH batteries fixed to the battery cover.
If the batteries do not hold a sufficient charge for
proper operation follow these steps for replacement
of the battery pack.
NOTE:
The battery pack fits in to the compartment
one way only. If the pack is inserted incorrectly the
battery cover will not close properly.
NOTE:
Only use battery packs supplied by the
manufacturer. The use of any other type of battery
will immediately invalidate the warranty!
NOTE:
Dispose the used battery pack in accordance of
your local regulations.
1. Hold the pipette in a horizontal position with the
display facing up.
2. Press the two charging contacts simultaneously to
release the battery pack.
3. Place the new battery pack into the battery
compartment and close the battery cover by
pressing it lightly.

11.4. Storage

When not in use it is recommended that the pipette
is always stored in the eLINE® Charging Stand or
Carousel. However, during long storage periods
(several months) it is recommended to disconnect the
charging stand and carousel from the main outlet.
It is also advisable to remove the battery pack from the
pipette. This protects the batteries against discharging
during the long storage period.

12. Performance Testing

It is recommended to check the performance of your
eLINE® pipettes regularly (e.g. every 3 months) and
always after in-house maintenance. However, users
should establish a regular testing routine for their
33