14. Cleaning / repairing the unit

Caution:
Always turn off the unit and disconnect the mains cable from the
power source before cleaning the unit.
Prevent humidity from entering into the circulator.
Electrical connections and any other work must be performed by
qualified personnel only.
To maintain the full cooling performance, clean the condenser
from time to time.
• Switch off the unit, disconnect mains power cable.
• Hold the venting grid, pull out and remove.
• Clean the ribbed condenser with a vacuum cleaner.
• Replace the venting grid.
• Switch on the unit.
Water cooled models:
In order to maintain a good condition of the cooling
compressor, the sieve in the cooling water input should be
cleaned in regular intervals.
Cleaning:
For cleaning the bath tank and the immersed parts of the
circulator, use low surface tension water (e.g., soap suds).
Clean the outside of the unit using a wet cloth and low
surface tension water.
The circulator is designed for continuous operation under
normal conditions. Periodic maintenance is not required.
The tank should be filled only with a bath fluid recommended
by JULABO. To avoid contamination, it is essential to change
the bath fluid from time to time.
Repairs
Before asking for a service technician or returning a
JULABO instrument for repair, please contact an
authorized JULABO service station.