Interface - Baud Rate, Handshake, Parity - Julabo F25-MC Operating Manual

Refrigerated and heating circulators
Hide thumbs Also See for F25-MC:
Table of Contents

Advertisement

Menu functions
9.5.

Interface - Baud rate, Handshake, Parity

Factory settings:
4.8 kBauds
2 = even
Hardware handshake
Example: 4.8 kbauds
Example: 2 = even
44
To allow communication of the circulator with a PC or a
superordinated process control system take care the
interface parameters of both units match.
Adjustable interface parameters
Baud rate
Parity
Handshake
0 = Xon/Xoff protocol
1 = Protocol RTS/CTS hardware handshake
Data bits = 7; Stop bits = 1
1.
Press the
displayed.
2.
Press enter
3.
Select the parameter with the keys
4.
Press enter
4.8 = 4800 bauds
9.6 = 9600 bauds
19.2 = 19200 bauds
38.4 = 38400 bauds
0 = no Parity
1 = odd
2 = even
software handshake
button until the desired menu item is
to indicate the parameter.
to store the selected parameter.
.