Safety Installations, Warning Functions; Excess Temperature Protection - Julabo F25-MC Operating Manual

Refrigerated and heating circulators
Hide thumbs Also See for F25-MC:
Table of Contents

Advertisement

8.

Safety installations, warning functions

Check the safety installations at least twice a year (see page 21)!
8.1.

Excess temperature protection

Warning:
The excess temperature protection >tSA< should be set at least 25 °C
below the fire point of the bath fluid used.
In the event of wrong setting there is a fire hazard!
We disclaim all liability for damage caused by wrong settings!
Settings for the excess temperature protection according to
IEC 61010-2-010 and for the high and low temperature
warning functions are made
in a menu that is called up with the key
This safety installation is independent of the control circuit.
When the temperature of the bath fluid has reached the safety
temperature, a complete shutdown of the heater and pump is
effected.
The alarm is indicated by optical and audible signals
(continuous tone) and on the MULTI-DISPLAY (LED) appears
the error message "Error 14".
1.
Press the key
2.
Press enter
The adjusted cut-out value is indicated on the MULTI-
DISPLAY (LED).
3.
Set the new cut-out value using a screwdriver via the
MULTI-DISPLAY (LED). (Example: 100 °C)
4.
Press
to update the display immediately.
Recommendation:
Set the excess temperature protection at 5 to 10 °C above
the working temperature setpoint.
and call up the menu
.
MC
.
.
33

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents