Safety installations, warning functions
8.3.

Limitation of the operating temperature range

Factory settings:
t HighSetpoint: 200 °C
t LowSetpoint: -50 °C
36
The warning functions are only triggered when the actual
bath temperature, after start from the „OFF" or „rOFF"
mode, lies within the set limits for 3 seconds.
Recommendation::
Set the high temperature warning value >t High< 5 °C
to10 °C above the working temperature setpoint
Set the low temperature warning value >t Low< 5 °C to
10 °C below the working temperature setpoint
The limitation of the operating temperature range effects the
temperature setting under the menu called up with the
key. It is possible to adjust only working temperatures that lie
within the limit range set here.
Existing settings for Setpoint 1, 2, 3 and also for > t High <
and > t Low < (see page 35) are automatically defered within
the limit range.
1.
Press the key
or
2.
Press enter
MULTI-DISPLAY (LED) (digit blinks).
3.
Set value:
Use the cursor keys
display until the numeral you wish to change blinks.
Use the increase/decrease arrows
selected numeral (-, 0, 1, 2, 3, ... 9).
4.
Press enter
Press
to update the display immediately.
until the menu item
is displayed.
to indicate the adjusted value on the
to move left or right on the
to store the selected value.
to change the