Status Messages; Error Messages - Julabo F25-MC Operating Manual

Refrigerated and heating circulators
Hide thumbs Also See for F25-MC:
Table of Contents

Advertisement

Remote control

12.4. Status messages

Status messages
00 MANUAL STOP
01 MANUAL START
02 REMOTE STOP
03 REMOTE START

12.5. Error messages

Error messages
-01 LOW LEVEL ALARM
-02 REFRIGERATOR ALARM
-03 EXCESS TEMPERATURE
WARNING
-04 LOW TEMPERATURE
WARNING
-05 WORKING SENSOR ALARM
-06 SENSOR DIFFERENCE
ALARM
2
-07 I
C-BUS ERROR
-08 INVALID COMMAND
-09 COMMAND NOT ALLOWED
IN CURRENT OPERATING
MODE
-10 VALUE TOO SMALL
-11 VALUE TOO LARGE
-12 TEMPERATURE
MEASUREMENT ALARM
-13 WARNING : VALUE
EXCEEDS TEMPERATURE
LIMITS
54
Description
Circulator in „OFF" state.
Circulator in keypad control mode.
Circulator in „r OFF" state.
Circulator in remote control mode.
Description
Low liquid level alarm.
Control cable of the refrigerated circulator or
MVS solenoid valve controller short-circuited or
interrupted.
High temperature warning.
Low temperature warning.
Working temperature sensor short-circuited or
interrupted.
Sensor difference alarm.
Working temperature and safety sensors report
a temperature difference of more than 25 °C.
Internal error when reading or writing the I
bus.
Invalid command.
Invalid command in current operating mode.
Entered value too small.
Entered value too large.
Error in A/D converter.
Value lies outside the adjusted range for the
high and low temperature warning limits. But
value is stored.
2