Setting The Pump Pressure - Julabo F25-MC Operating Manual

Refrigerated and heating circulators
Hide thumbs Also See for F25-MC:
Table of Contents

Advertisement

9.1.

Setting the pump pressure

Factory setting:
Stage 2
B
internal100 %
C
internal 20 %
external 80 %
The pressure of the circulating pump is adjustable in four
grades
Flow rate:
Pump pressure:
1.
Press the
2.
Press enter
3.
Select the parameter with the keys
4.
Press enter
Press
Adjusting the pump flow
The pump flow is pre-adjusted in the factory and can be
2
modified to suit user requirements.
• Using a screwdriver turn the screw (1) anti-clockwise
by 360 °.
1
• Using flat pliers turn the marking of the slide (2) to the
desired position.
• Tighten the screw.
Examples:
A
Internal applications in the bath
A 100 % internal bath circulation
(for large bath tanks)
B
Reduced internal bath circulation
(for smooth surface of bath fluid)
External/internal applications
C 40 % external discharge,
60 % internal circulation
(for large bath tanks)
D 80 % external discharge,
D
20 % internal circulation
(for small bath tanks)
11 ... 16 l/min
0.22 ... 0.45 bar
button until the menu item
is displayed.
to indicate the parameter.
to store the selected parameter.
to update the display immediately.
MC
(1 ... 4).
39

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents